Medtronic Valleylab FT10

Error: E323 Check Foot Pedal

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

Medtronic

Model

Valleylab FT10

What This Guide Helps With

This guide helps troubleshoot Error: E323 Check Foot Pedal on the Medtronic Valleylab FT10 Electrosurgical Unit. This error indicates the Mono 2 foot pedal may be pressed, stuck, connected incorrectly, damaged, or falsely detected as active.

A foot pedal activation-control issue can prevent normal ESU readiness, interrupt use, or create an unintended activation risk if the pedal input is not reliable.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Release and Inspect the Mono 2 Foot Pedal

Check the Foot Pedal Connection

Confirm the Correct Foot Pedal and Port

Isolate the Foot Pedal

Swap with a Known-Good Foot Pedal

Inspect for Cable Damage or Intermittent Faults

Check for Environmental or Setup Causes

Determine Whether the Fault Is Accessory-Related or Device-Related

If the Problem Persists

If Error E323 continues after releasing the pedal, checking the connector, confirming correct setup, removing the pedal, checking the environment, and testing with a known-good compatible foot pedal, external and common causes have been ruled out.

Persistent E323 may indicate an internal ESU detection problem involving the foot pedal input path, Controller PCBA, Steering Relay PCBA, Foot Pedal PCBA, or related circuitry.

Clinical Use Tip

Never troubleshoot a foot pedal activation error on an active patient. A foot pedal that is stuck, shorted, or falsely detected can create an unintended activation risk. Move the patient to a backup device first, then continue troubleshooting in a safe, non-patient setup.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Medtronic Valleylab FT10 displayed Error E323 Check Foot Pedal. Staff reported Mono 2 foot pedal alarm during setup."

Cause

What was observed during troubleshooting.

Example:
"Original Mono 2 foot pedal remained detected as pressed. Error cleared when pedal was removed and did not return with known-good compatible foot pedal."

Resolution

What action was taken.

Example:
"Removed defective foot pedal from service. Installed known-good compatible foot pedal, verified normal startup and safe pedal response. Returned ESU to service."

Helpful Details to Include (If Known)

Final Thought

Error E323 should be treated as a foot pedal activation-control problem until proven otherwise. Start with patient safety, confirm the exact alarm, check the pedal and connection, then isolate with a known-good foot pedal. If the error remains after external causes are ruled out, remove the ESU from service and escalate for repair. Clear CCR documentation protects patient safety and shows exactly how the fault was isolated.

That is successful troubleshooting.