Medtronic Valleylab FT10

Error: E355 Invalid Activation

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

Medtronic

Model

Valleylab FT10

What This Guide Helps With

Error E355 Invalid Activation indicates that the Valleylab FT10 has detected an unsupported activation request, commonly when a foot pedal is used with an instrument or setup that requires activation from the instrument only. This can prevent proper ESU output and delay cutting, coagulation, bipolar, or vessel sealing workflow.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Confirm Correct Activation Method

Dismiss the Error and Reactivate Correctly

Inspect the Foot Pedal Connection

Try Operation Without the Foot Pedal

Swap the Foot Pedal if Clinically Appropriate

Check for Unintended Foot Pedal Activation

Check Instrument and Setup

Isolate Accessory vs Device Issue

If the Problem Persists

If E355 continues after removing or replacing the foot pedal, confirming the correct activation method, checking the instrument, and using known-good compatible accessories, external causes have been ruled out.

Clinical Use Tip

Never troubleshoot activation errors on an active patient. Move the patient to a backup device first if electrosurgical energy is needed. If the FT10 reports E355, confirm whether the connected instrument supports foot pedal activation before assuming the generator has failed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"User reported Medtronic Valleylab FT10 displaying Error E355 Invalid Activation during attempted foot pedal activation."

Cause

What was observed during troubleshooting.

Example:
"Foot pedal activation was not supported for the connected instrument/setup. Error cleared when activation was performed from the instrument controls."

Resolution

What action was taken.

Example:
"Dismissed error, verified correct activation method, inspected foot pedal and instrument connections, confirmed normal operation using instrument activation only, and returned ESU to service."

Helpful Details to Include (If Known)

Final Thought

Error E355 should be approached as an activation-method problem first, not an immediate internal failure. Start with patient safety, confirm the exact error, verify whether the instrument supports foot pedal activation, and rule out damaged accessories before escalating. If the message persists with correct setup and known-good accessories, remove the FT10 from service for further evaluation.

That is successful troubleshooting.