On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
A startup message, logged event, or service indication remains present after normal corrective actions or returns after the generator is restarted.
Step-by-Step Troubleshooting
1. Protect the Patient
If the message affects generator readiness, energy delivery, alarms, or any safety function, remove the generator from clinical use and provide a verified replacement.
Do not clear or suppress a message merely to continue a procedure.
Expected outcome: Patient care does not depend on a generator with an unresolved fault indication.
2. Record the Exact Message
Document the complete displayed wording, code, sequence, and circumstances before restarting or changing anything.
Do not rely on a verbal description such as “service error” if more specific information is available.
Expected outcome: The reported condition can be accurately referenced during troubleshooting or escalation.
3. Determine Whether the Message Is Active or Historical
Establish whether the generator currently has an active fault or whether the information exists only as a stored event.
Do not assume a historical event log entry requires repair if the generator currently passes all required checks.
Expected outcome: Active faults are distinguished from historical records.
4. Check Power and External Accessories
Verify reliable AC power and disconnect nonessential accessories that could be contributing to startup or recognition faults.
Restart the generator normally.
Expected outcome: The message clears if it was caused by an external accessory or transient power condition. If it remains cleared and full verification passes, troubleshooting can stop.
5. Reconnect Accessories Individually
Reconnect approved accessories one at a time and observe whether the error returns.
Inspect each associated cable and connector before reconnecting.
Expected outcome: Any external accessory associated with recurrence is identified.
6. Check for Environmental or Physical Causes
Inspect for blocked ventilation, fluid contamination, damaged connectors, impact damage, or other external conditions that could plausibly generate the fault.
Expected outcome: No obvious external physical or environmental problem remains.
7. Use Only Authorized Log or Service Functions
Review available diagnostic information only through interfaces and procedures authorized for Clinical Engineering use.
Do not enter unauthorized service modes, alter protected configuration values, or repeatedly clear codes without identifying the cause.
Expected outcome: Diagnostic information is preserved and used appropriately without masking the underlying fault.
8. Perform Functional Verification
If the active fault clears after correcting an external cause, verify startup, controls, alarms, accessory recognition, REM function, and applicable ESU outputs using approved test equipment.
Expected outcome: The generator passes required verification and the error does not recur. Troubleshooting can stop.
9. Escalate Recurring or Uncleared Errors
If an active startup error or service indication repeatedly returns despite known-good power and external accessories, remove the generator from service.
Expected outcome: The underlying service-level issue is investigated rather than repeatedly erased.
If the Problem Persists
Common external causes have been eliminated. Persistent or recurring active codes may involve software, configuration, sensing, power, RF, safety-monitoring, or other internal service-level systems.
The Valleylab FX8 should be:
- Removed from service when the indication represents an active fault.
- Labeled Out of Service.
- Sent for repair or bench evaluation.
- Evaluated using appropriate Medtronic documentation and approved diagnostic tools.
- Repaired, configured, or software-serviced only by qualified personnel.
After corrective service, verify that the fault does not recur and complete all required return-to-service testing.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Preserve the exact error information before clearing anything; the original message may be the most useful evidence for service evaluation.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported that a startup service message repeatedly returned after the Valleylab FX8 was restarted."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the error recurred only when a damaged external accessory was connected to the generator."
Resolution
What action was taken.
Example:
"The defective accessory was removed from service, and the generator completed repeated startup, alarm, accessory, and ESU functional verification without recurrence."
Helpful Details to Include (If Known)
- Exact displayed message or code
- Whether code is active or historical
- Time or operating condition when it occurred
- Power source
- Accessories connected
- Results with accessories disconnected
- Event-log observations
- Whether the message recurs after restart
- Functional-test results
- Final device status
Final Thought
Do not treat clearing a code as repairing a fault; document the indication, identify external causes first, verify full function, and escalate recurring active errors appropriately.
That is successful troubleshooting.