Midmark Ritter 75L / 75LE Series

Backrest Not Adjusting or Drifting

On this page

Asset Type

Exam Table

Manufacturer

Midmark

Model

Ritter 75L / 75LE Series

What This Guide Helps With

Backrest not moving, slow movement, or drifting down due to control, power, or external mechanical causes.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Remove the table from active patient use before troubleshooting.

Verify the Reported Issue

Operate the backrest using the hand control or foot control. Observe for no movement, delayed movement, or drifting after release.

Check Power to the Table

Confirm the table is plugged into a working outlet. Test outlet with another device.

Inspect Hand/Foot Control Function

Check all buttons on the control. Confirm other functions (hi-lo) operate normally.

Inspect Control Cable and Connections

Check the hand/foot control cable for damage, pinching, or loose connections at the table base.

Check for Obstruction or Binding

Inspect the backrest hinge area and surrounding frame for debris, misalignment, or mechanical interference.

Assess Load and Weight Distribution

Ensure weight is within limits and evenly distributed on the backrest.

Listen for Actuator Operation

Operate the backrest and listen for motor noise.

Check for Backrest Drift After Movement Stops

Raise the backrest and release the control. Observe if it slowly lowers.

If the Problem Persists

If all external causes (power, controls, cables, obstructions, and loading) have been ruled out, the issue is likely internal to the actuator or control system.

Remove the table from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and ensures proper repair.

Clinical Use Tip

Never troubleshoot backrest movement with a patient on the table. Always transfer the patient first to prevent injury from unexpected movement.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Backrest not adjusting and drifting down during use."

Cause

What was observed during troubleshooting.

Example:
"Backrest actuator unable to hold position after ruling out power, control, and mechanical obstruction issues."

Resolution

What action was taken.

Example:
"Removed from service and sent for actuator replacement/repair."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with simple external checks and progresses logically. Prioritizing patient safety and recognizing likely actuator failure helps avoid unnecessary downtime and risk. Proper documentation ensures clear communication and supports efficient repair.

That is successful troubleshooting.