Mindray BeneVision N Series

Alarm: CMS/eGW Disconnected

On this page

Asset Type

Patient Monitor

Manufacturer

Mindray

Model

BeneVision N Series

What This Guide Helps With

This guide helps troubleshoot the CMS/eGW Disconnected alarm on the Mindray BeneVision N Series patient monitor. This alarm means the bedside monitor has lost communication with the central monitoring system or eGateway. The monitor may still function locally at the bedside, but patient data may not be visible at the central station.

This can affect central monitoring visibility, alarm awareness, patient tracking, and clinical workflow. The goal is to confirm patient safety first, then check the most common external causes such as network cable connection, wall jack, monitor assignment, and central station communication.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm the Bedside Monitor Is Working Locally

Check the Network Cable Connection

Check the Network Wall Jack or Port

Replace the Network Cable With a Known-Good Cable

Confirm Central Station Visibility

Check for Recent Monitor Movement or Bed Assignment Changes

Restart the Monitor Only If Clinically Safe

Isolate Monitor, Dock, Room, or Network Cause

Check for a Broader CMS/eGW or Network Issue

If the Problem Persists

If the CMS/eGW Disconnected alarm remains after checking the network cable, known-good cable, wall jack, monitor assignment, central station visibility, restart, and alternate network location, external and common causes have been ruled out.

Persistent disconnection may indicate a faulty network path, bad wall jack, incorrect network configuration, docking station communication issue, monitor network interface issue, or CMS/eGateway communication problem.

Remove the monitor from central-monitoring-dependent use if reliable CMS/eGW communication cannot be restored. Label the device Out of Service if the issue appears monitor-related. Escalate to Clinical Engineering, IT/network support, central monitoring support, or the vendor as appropriate. Send the monitor or dock for repair or bench evaluation if the problem follows the device.

Clinical Use Tip

Never troubleshoot this alarm while assuming the central station is still seeing the patient. Verify local bedside monitoring first. If central monitoring is required and the connection cannot be restored quickly, move the patient to a backup monitor or verified monitored bed before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Mindray BeneVision N Series patient monitor displayed CMS/eGW Disconnected alarm. Staff reported the patient was not visible at the central station."

Cause

What was observed during troubleshooting.

Example:
"Monitor continued local bedside monitoring, but CMS/eGW communication was lost. Network cable, wall jack, monitor assignment, and central station visibility were checked."

Resolution

What action was taken.

Example:
"Verified patient remained monitored locally. Reseated network cable, replaced Ethernet cable with known-good cable, and confirmed live waveforms and numerics were restored at the central station. Monitor returned to service."

Helpful Details to Include (If Known)

Final Thought

A CMS/eGW Disconnected alarm should be handled as a patient visibility and communication issue first. Protect the patient, verify bedside monitoring, check simple network causes, confirm live central station data, and escalate when the issue points beyond the bedside setup. Clear documentation helps separate cable, room network, configuration, dock, monitor, and CMS/eGateway failures.

That is successful troubleshooting.