Mindray BeneVision N Series

Alarm: Power Board Comm Error

On this page

Asset Type

Patient Monitor

Manufacturer

Mindray

Model

BeneVision N Series

What This Guide Helps With

This guide helps troubleshoot the Power Board Comm Error alarm on a Mindray BeneVision N Series patient monitor. This alarm means communication with the power board has failed. The condition may affect AC power detection, battery charging, battery status reporting, power stability, or continued monitor operation.

Because this is a power-related communication alarm, the monitor should not be trusted for active patient monitoring until patient safety is confirmed and the device is verified.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check the AC Power Source

Check the Power Cord and Monitor Power Connection

Check Battery Status and Basic Power Behavior

Restart the Monitor

Inspect for Physical Damage or Environmental Causes

Isolate Docking or External Power Setup Issues

Perform Basic Functional Verification

Determine Whether the Fault Is Internal

If the Problem Persists

If the Power Board Comm Error persists after restart, AC power checks, power cord inspection, battery checks, and docking or external setup checks, the common external causes have been ruled out.

A persistent alarm may indicate an internal fault involving the power board, main control board, internal communication path, or related internal power circuitry.

Clinical Use Tip

Never troubleshoot this alarm on an active patient if the monitor is needed for continuous monitoring. Move the patient to a backup monitor first. A power communication fault can affect monitor reliability even if the screen is currently on.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Mindray BeneVision N Series patient monitor displays Power Board Comm Error during startup and may not reliably report power or battery status."

Cause

What was observed during troubleshooting.

Example:
"External AC outlet, power cord, battery status, and docking setup were checked. Alarm persisted after controlled restart, indicating likely internal communication failure between the main control system and power board."

Resolution

What action was taken.

Example:
"Removed monitor from patient use, labeled Out of Service, and sent for bench repair evaluation. Power board or main control board replacement may be required depending on service findings."

Helpful Details to Include (If Known)

Final Thought

A Power Board Comm Error should be handled as a reliability issue, not just a nuisance alarm. Start with patient safety, verify the exact alarm, rule out external power and connection problems, then escalate when the alarm persists. Proper documentation shows that troubleshooting was logical and that the device was removed from service before becoming a patient safety risk.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?