Mindray BeneVision N Series

Alarm: RT CLOCK NEED RESET

On this page

Asset Type

Patient Monitor

Manufacturer

Mindray

Model

BeneVision N Series

What This Guide Helps With

This guide helps troubleshoot the RT CLOCK NEED RESET alarm on a Mindray BeneVision N Series patient monitor. This alarm means the monitor real-time clock needs to be reset. It may affect time stamps, trends, alarm history, event review, and documentation accuracy.

The condition may be caused by an incorrect date or time setting, a clock reset after power loss, depleted backup power, a weak button battery, or a main control board issue.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm

3. Reset the Date and Time

4. Confirm the Monitor Retains Time After Restart

5. Check External Power Conditions

6. Inspect for Physical or Environmental Issues

7. Identify the Likely Cause

If the Problem Persists

If the RT CLOCK NEED RESET alarm returns after the time is reset, the monitor is restarted, external power is verified, and obvious external causes are ruled out, the issue is no longer a simple clock setting problem.

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. The monitor may require clock backup battery service, main control board evaluation, or vendor repair depending on facility policy and supported service procedures.

Clinical Use Tip

Never troubleshoot this alarm on an active patient when accurate timing is important for care, documentation, alarm review, or trend review. Move the patient to a backup monitor first, then troubleshoot the affected device away from patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Mindray BeneVision N Series patient monitor displayed RT CLOCK NEED RESET alarm during startup and the monitor date/time required reset."

Cause

What was observed during troubleshooting.

Example:
"Real-time clock setting was incorrect. Date and time were reset, monitor was restarted, and clock retention was checked. If alarm returns, suspected clock backup battery or main control board issue."

Resolution

What action was taken.

Example:
"Reset monitor date and time, restarted monitor, and verified the correct time was retained. Alarm cleared and monitor passed normal startup checks. Returned to service."

Helpful Details to Include (If Known)

Final Thought

The RT CLOCK NEED RESET alarm should be handled logically. Protect patient safety first, verify the exact alarm, reset the clock, confirm the monitor can retain the corrected time, and rule out external power or storage conditions. If the alarm returns after reset and restart, remove the monitor from service and escalate it for proper repair evaluation. Clear CCR documentation shows what was reported, what was checked, and why the device was returned to service or removed from use.

That is successful troubleshooting.