Mindray BeneVision N Series

Alarm: THE PATIENT DATA STORAGE SPACE IS NEARLY FULL. PLEASE DELETE SOME DISCHARGED PATIENTS.

On this page

Asset Type

Patient Monitor

Manufacturer

Mindray

Model

BeneVision N Series

What This Guide Helps With

This guide helps troubleshoot the Mindray BeneVision N Series alarm indicating that patient data storage space is nearly full and unnecessary discharged patient records need to be deleted. This condition usually does not mean monitoring has failed, but it can affect patient data management, trend storage, admission workflow, and future record handling if not corrected.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm the Monitor Is Not Being Used for a Patient Transfer or Documentation Review

Delete Unnecessary Earlier Discharged Patients

Recheck the Monitor After Deletion

Check for Repeated Storage Alarms

Determine Whether the Issue Is Workflow-Related or Device-Related

If the Problem Persists

If discharged patient records have been deleted and the alarm still remains, external and common workflow causes have been ruled out.

A persistent storage alarm may indicate a software, database, or internal memory/storage problem.

The monitor should be removed from patient use, labeled Out of Service, and sent for repair, bench evaluation, or vendor/service support.

Do not continue using a patient monitor that cannot reliably store or manage patient data.

Clinical Use Tip

Never troubleshoot on an active patient in a way that risks monitoring interruption or loss of needed patient information. Move the patient to a backup monitor first if the alarm affects active monitoring, admission workflow, trend storage, or patient data reliability. Confirm with clinical staff before deleting discharged patient records.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Mindray BeneVision N Series patient monitor displayed alarm: THE PATIENT DATA STORAGE SPACE IS NEARLY FULL. PLEASE DELETE SOME DISCHARGED PATIENTS. Staff reported the alarm appeared during normal monitoring or admission workflow."

Cause

What was observed during troubleshooting.

Example:
"Patient data storage space was nearly full due to accumulated earlier discharged patient records. Current patient monitoring remained active and no external accessory, parameter, power, or network issue was found."

Resolution

What action was taken.

Example:
"Confirmed active patient monitoring was stable. Verified alarm message. With clinical staff approval, unnecessary discharged patient records were deleted. Alarm cleared, current patient data remained correct, and monitor was returned to service."

Helpful Details to Include (If Known)

Final Thought

This alarm should be approached logically. Confirm the message, protect active patient monitoring, verify that discharged patient data can be safely deleted, and clear unnecessary records. If the alarm persists after normal storage cleanup, escalate the monitor for service rather than forcing it back into use.

That is successful troubleshooting.