Mindray SynoVent E5

Pre-use check or leak test failure

On this page

Asset Type

Ventilator

Manufacturer

Mindray

Model

SynoVent E5

What This Guide Helps With

This guide assists in troubleshooting pre-use check or leak test failures on a Mindray SynoVent E5 ventilator. These failures are often caused by external setup issues such as breathing circuit leaks, misconnected components, or sensor problems, rather than internal ventilator faults. Following a systematic approach ensures patient safety and avoids unnecessary internal interventions.

Step-by-Step Troubleshooting

Verify Power and Connections

Check Breathing Circuit Setup

Inspect Circuit for Leaks

Check Exhalation Valve and Patient Interface

Run Self-Test in Isolation

Inspect Sensors and Flow Modules

Review Software/Settings

If the Problem Persists

External causes have been ruled out; the failure is likely internal.

Remove the ventilator from service.

Label as “Out of Service.”

Send to Biomedical Engineering bench evaluation or manufacturer service.

Knowing when to escalate is proper troubleshooting.

Clinical Use Tip

Never perform troubleshooting on an active patient.

Move the patient to a functioning ventilator before testing.

Ensure all alarms are active and the device is in a safe state before attempting pre-use checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
User reports pre-use check fails with “Leak Test Failure” alarm on SynoVent E5.

Cause

What was observed during troubleshooting.

Example:
Visual inspection revealed loose elbow connector in breathing circuit; all sensors properly seated.

Resolution

What action was taken.

Example:
Reassembled circuit, retested, pre-use check completed successfully. Device returned to service.

Helpful Details to Include (If Known)

Final Thought

Systematic troubleshooting ensures patient safety, avoids unnecessary internal interventions, and maintains equipment reliability. Thorough documentation provides clear communication for future evaluations and reinforces clinical engineering best practices.

That is successful troubleshooting.