On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots image degradation isolated to one transducer by checking coupling, probe condition, settings, connections, interference, and known-good comparisons.
Step-by-Step Troubleshooting
1. Protect the Patient From Unreliable Imaging
Do not make clinical decisions from an image known to contain unexplained dropout, noise, or nonuniformity. Switch to a verified transducer or ultrasound system.
Expected outcome: Patient imaging continues with a reliable image source.
2. Confirm the Artifact
Determine whether the artifact is consistent, intermittent, localized, depth-dependent, or associated with movement of the transducer cable or connector.
Expected outcome: The artifact is reproducible under controlled nonclinical testing.
3. Verify Coupling and Test Conditions
Use an appropriate approved test method and ensure adequate acoustic coupling. Eliminate air gaps, insufficient gel, contaminated probe surfaces, or poor contact as causes.
Expected outcome: A properly coupled probe produces a consistent test image. If correcting coupling eliminates the artifact, verify and stop.
4. Inspect the Transducer
Inspect the probe face, housing, cable, strain relief, and connector for cuts, separation, impact damage, delamination, contamination, or fluid intrusion.
Expected outcome: No visible damage is present. A damaged transducer is removed from service.
5. Reseat the Transducer Connection
Disconnect and reconnect the probe appropriately, ensuring the connector is fully seated and secured.
Expected outcome: The transducer is recognized reliably and the image remains stable.
6. Return to a Known Baseline Preset
Use an appropriate standard exam preset or known baseline user configuration to rule out excessive gain, processing, depth, focus, or other user-controlled image settings.
Expected outcome: Normal image uniformity returns if inappropriate settings caused the appearance.
7. Check for Environmental Electrical Noise
Observe whether the artifact changes when nearby nonessential electrical equipment is turned off or separated where appropriate. Inspect the environment for obvious sources of interference without defeating grounding or safety systems.
Expected outcome: The image remains free of unexpected noise under normal clinical environmental conditions.
8. Compare With a Known-Good Transducer
Test a compatible known-good transducer using the same system and equivalent test conditions. If possible, test the suspect probe on another compatible system.
Expected outcome: The comparison determines whether the artifact follows the transducer or remains with the TE X.
9. Perform Final Image-Quality Verification
Using appropriate ultrasound quality-assurance methods and approved test equipment, verify image uniformity and absence of abnormal dropout or noise.
Expected outcome: Image quality is suitable for clinical use. Persistent unexplained artifacts require removal of the affected probe or system from service.
If the Problem Persists
Coupling, probe condition, connector seating, image settings, environmental interference, and known-good comparisons have been addressed. Remaining causes may involve transducer elements, cable conductors, connector interfaces, receive electronics, image processing, or another internal condition.
Remove the affected transducer or TE X from service based on the comparison results. Label it Out of Service and obtain repair or bench evaluation using appropriate Mindray documentation and approved ultrasound test equipment.
Do not attempt internal transducer or board repair. Complete appropriate image-quality and functional tests before return to clinical service.
Clinical Use Tip
When an artifact appears on only one probe, verify the probe pathway before interpreting the pattern as anatomy or pathology.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported a persistent dark dropout region when using one transducer on the TE X."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found visible damage to the suspect transducer cable, and the artifact followed that transducer during comparison testing."
Resolution
What action was taken.
Example:
"Clinical Engineering removed the damaged transducer from service and verified normal uniform imaging on the TE X with a known-good compatible transducer."
Helpful Details to Include (If Known)
- Transducer identifier
- Type and location of artifact
- Probe-face and cable condition
- Coupling method
- Preset used
- Connector and port tested
- Known-good comparison results
- Environmental interference observations
- Final image-quality result
- Final probe and system status
Final Thought
Treat unexplained image degradation as a patient-safety concern, establish whether the problem follows the probe or system, and verify image quality before anything returns to clinical use.
That is successful troubleshooting.