On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots external display or printer detection failures caused by power, cables, input selection, connection, compatibility, configuration, or peripheral faults.
Step-by-Step Troubleshooting
1. Protect the Clinical Workflow
If the external display or printer is required for an active procedure or required documentation, establish an approved alternate method before troubleshooting.
Expected outcome: Clinical care and required documentation continue without dependence on the failed peripheral.
2. Confirm Which External Function Has Failed
Determine whether the TE X fails to detect an external monitor, produces no video despite detection, cannot communicate with a printer, or loses the peripheral intermittently.
Expected outcome: The issue is narrowed to detection, signal output, printing, or intermittent communication.
3. Verify Peripheral Power
Confirm the monitor or printer is powered, shows normal status, and is connected to a known-good power source.
Expected outcome: The peripheral powers normally. If restoring peripheral power resolves the issue, complete functional verification and stop.
4. Inspect External Cables and Connectors
Inspect the complete accessible video or printer cable for damage, loose connectors, bent or damaged contacts, adapters, or strain. Reseat connections securely.
Expected outcome: The cable and connectors are secure and undamaged. If reseating restores normal operation, verify stability and stop.
5. Verify External Monitor Input
For a monitor issue, confirm the display is set to the physical input connected to the TE X and that no other source is selected.
Expected outcome: The external monitor displays the TE X output normally when the correct input is selected.
6. Verify User-Accessible Output or Printer Selection
Using normal TE X controls, confirm the intended external output or approved printer is selected and enabled. Do not alter restricted service configuration.
Expected outcome: The intended external peripheral is recognized and available.
7. Substitute Known-Good External Components
When possible, test with a known-good compatible cable, display, or approved printer. Alternatively, test the suspect peripheral with a known-good source when appropriate.
Expected outcome: The comparison identifies whether the failure follows the peripheral/cable or remains with the TE X.
8. Check Connection Stability
Observe the peripheral during normal cart and cable positioning. Do not force connectors. Determine whether ordinary movement causes video loss or peripheral disconnection.
Expected outcome: Detection and output remain stable throughout normal positioning.
9. Perform Final Functional Verification
For video, verify a stable image on the external monitor through the expected workflow. For printing, produce an approved test output and confirm it is complete and legible.
Expected outcome: The peripheral is detected and performs its intended function reliably. If detection remains intermittent or absent, escalate.
If the Problem Persists
Peripheral power, cable condition, connector seating, display input selection, user-accessible settings, and known-good substitutions have been checked. Remaining causes may involve the TE X output interface, software, peripheral compatibility, configuration, internal communication hardware, or the peripheral itself.
Remove the TE X from service if the external function is required for safe clinical use and no approved alternative exists. If the failure is isolated to a nonessential peripheral, clearly document the restriction and follow institutional policy.
Label failed equipment Out of Service as appropriate and send it for repair or bench evaluation using appropriate Mindray and peripheral documentation and approved test equipment. Complete the relevant output or printing workflow before return to service.
Clinical Use Tip
For procedure displays, verify the actual external image before patient use rather than assuming that a connected cable means the video path is functional.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the external procedure monitor connected to the TE X remained blank during setup."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the external monitor was configured for a different video input than the connection used by the TE X."
Resolution
What action was taken.
Example:
"Clinical Engineering selected the correct monitor input and verified stable TE X video output during functional testing and normal cart positioning."
Helpful Details to Include (If Known)
- Peripheral type
- Peripheral power status
- Cable and adapter condition
- Connection used
- Monitor input selected
- TE X output selection
- Known-good cable/peripheral results
- Whether movement caused interruption
- Test print or video result
- Final device status
Final Thought
Verify the external peripheral, cable, input, and normal configuration before suspecting an internal output failure, then confirm the complete clinical workflow and escalate unresolved or intermittent communication problems.
That is successful troubleshooting.