Mindray WATO EX Series

Bellows Not Rising or Cycling

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

What This Guide Helps With

Troubleshooting absent, incomplete, or irregular bellows movement caused by ventilation settings, circuit leaks, breathing-system assembly, gas supply, or external components.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended bellows or ventilator troubleshooting while the WATO EX Series is supporting an active patient.

If the bellows stop rising or cycling during clinical use:

Expected outcome: Patient care no longer depends on an anesthesia machine with uncertain ventilation.

Continue troubleshooting only after the machine has been removed from active patient use.

2. Confirm the Exact Failure Condition

Determine whether:

Record the complete alarm or advisory message before restarting the machine.

Expected outcome: The failure pattern is clearly identified and can guide the remaining checks.

3. Verify Ventilation Mode and Basic Settings

Confirm that:

Use a manufacturer-approved test lung instead of a patient.

Expected outcome: The ventilator initiates breaths and the bellows begin cycling.

If normal operation returns, complete the required system checkout and stop troubleshooting.

4. Check the Bag/Ventilator Selection Control

Confirm that the bag/ventilator selector is fully positioned in the ventilator setting.

Move the selector through its normal range and return it firmly to the ventilator position. Do not force the control.

A selector left between positions may prevent proper ventilator operation or redirect breathing-system flow.

Expected outcome: Mechanical ventilation engages and the bellows cycle normally.

5. Verify Fresh-Gas and Drive-Gas Supplies

Check that:

The WATO EX family is available in several configurations, so gas sources and installed features may vary by model.

Expected outcome: Adequate gas supply is available and gas-related alarms clear.

If the bellows begin operating after restoring the gas supply, perform the complete pre-use checkout before returning the unit to service.

6. Inspect the Patient Breathing Circuit

Inspect the entire circuit from the inspiratory and expiratory ports to the test lung.

Look for:

Reconnect or replace questionable external components with compatible, approved accessories.

Expected outcome: The breathing circuit is complete, unobstructed, and capable of holding pressure.

7. Check for a Major Breathing-System Leak

Occlude the patient connection using the approved test method and perform the machine’s system checkout or leak test.

Pay particular attention to:

Do not continue mechanical ventilation testing if the breathing system cannot pass the required leak test.

Expected outcome: The system passes its leak test and the bellows rise normally between breaths.

8. Inspect the Bellows Assembly Externally

With the machine powered off and removed from service, visually inspect the accessible bellows assembly.

Confirm that:

Do not disassemble internal ventilator or drive-gas components unless authorized by the applicable Mindray service documentation.

Expected outcome: The bellows and housing are visibly intact, correctly installed, and free from obstruction.

9. Check the Absorber and Breathing-System Assembly

Confirm that:

Incorrect seating or assembly can create a substantial leak or restrict gas movement.

Expected outcome: The breathing system is assembled correctly and the bellows fill and cycle normally.

10. Inspect Expiratory Flow and Exhaust Paths

Check the accessible expiratory path for:

Temporarily substitute known-good compatible external tubing or filters when appropriate.

Do not operate the machine clinically with the scavenging system disconnected unless specifically permitted by facility procedures and the manufacturer’s instructions.

Expected outcome: Gas can exit the breathing system normally without excessive restriction or suction.

11. Run the Automated System Checkout

After correcting external issues:

Expected outcome: The machine completes checkout, delivers consistent breaths, and the bellows rise and compress normally.

Do not return the unit to service solely because the bellows move. All required checkout tests and performance verification must pass.

If the Problem Persists

If the bellows still do not rise or cycle after settings, gas supplies, circuit connections, leaks, breathing-system assembly, valves, absorber seating, and external restrictions have been checked, common external causes have been ruled out.

The problem may involve an internal ventilator valve, drive-gas circuit, pneumatic control component, pressure sensor, flow sensor, electronic control system, or internal seal.

The device should be:

Knowing when to stop and escalate is proper troubleshooting. Do not continue internal disassembly without appropriate training, authorization, and service information.

Clinical Use Tip

Never troubleshoot unreliable bellows movement on an active patient. Move the patient to another verified anesthesia machine or approved ventilation method before disconnecting circuits, testing gas supplies, or running checkout procedures.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the WATO EX Series bellows did not rise and the ventilator was not cycling during setup."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the absorber assembly incompletely seated, creating a breathing-system leak that prevented the bellows from filling."

Resolution

What action was taken.

Example:
"Reseated and secured the absorber assembly, completed the system leak test and automated checkout, verified consistent ventilation with a test lung, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Patient safety comes first whenever bellows movement or ventilation is unreliable. A logical evaluation of settings, gas supplies, circuits, leaks, breathing-system assembly, and external restrictions can resolve many failures without unnecessary disassembly. When the machine cannot pass checkout, remove it from service, escalate appropriately, and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide