On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting failed flow-sensor calibration caused by moisture, contamination, incorrect installation, breathing-system configuration, leaks, or unsuitable calibration conditions.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not perform flow-sensor calibration, remove breathing-system components, or disturb pneumatic connections while the WATO EX Series is supporting an active patient.
If the failure occurs during clinical use:
- Notify the anesthesia provider immediately.
- Confirm ventilation using airway pressure, capnography, oxygen saturation, chest movement, and an independent volume measurement when available.
- Move the patient to another verified anesthesia machine or approved ventilation method if delivered or measured volumes are questionable.
- Maintain appropriate independent patient monitoring.
Expected outcome: Patient care no longer depends on equipment with uncertain flow or volume measurement.
Continue troubleshooting only after the machine has been safely removed from patient use. Mindray specifies that flow-sensor calibration must not be performed while the unit is connected to a patient.
2. Confirm the Exact Failure Condition
Record the complete displayed message and determine whether the issue involves:
- Calibration that will not begin
- Calibration that begins but fails
- Inspiratory or expiratory flow readings remaining unavailable
- Abnormal tidal-volume readings after calibration
- A failure appearing after cleaning or breathing-system replacement
Expected outcome: The failure is clearly documented and can be reproduced safely.
If the calibration completes successfully and the system checkout passes, stop troubleshooting.
3. Allow the Machine to Complete Startup
Verify that:
- The machine has completed its startup sequence.
- No other checkout, leak-test, gas-supply, or ventilator faults are active.
- Pipeline gas supplies are connected and within the facility’s normal operating range.
- The machine is not being ventilated manually or mechanically during calibration.
Do not operate pneumatic controls, adjust gas flows, or disturb the breathing system while calibration is running.
Expected outcome: The machine is stable and ready to perform an uninterrupted calibration.
If calibration now passes, document the result and stop.
4. Inspect the Breathing System Externally
Confirm that the breathing-system assembly is:
- Fully seated on the anesthesia machine
- Correctly latched or secured
- Free from visible gaps or misalignment
- The correct assembly for that specific machine
- Not recently exchanged with an assembly from another WATO unit
Some WATO EX configurations are calibrated to their associated breathing-system block. Replacing or exchanging that assembly may require flow-sensor recalibration.
Expected outcome: The correct breathing system is properly installed without visible mechanical misalignment.
If reseating the assembly resolves the calibration failure, complete the system checkout and stop.
5. Inspect the Flow Sensors
Following facility procedures and the applicable Mindray instructions, remove and inspect the accessible inspiratory and expiratory flow sensors.
Check for:
- Condensation or water droplets
- Residual cleaning solution
- Secretions or visible contamination
- Cracked sensor housings
- Damaged sealing surfaces
- Missing, damaged, or displaced seals
- Bent or damaged connection points
- Incorrect orientation
Do not insert tools into the sensor, scrape sensing elements, or use unapproved cleaning methods.
Expected outcome: Both sensors are clean, dry, undamaged, and suitable for reinstallation.
If contamination or moisture is found, clean and dry the components only according to the approved procedure before continuing.
6. Verify Correct Sensor Installation
Reinstall each sensor in its designated location and orientation.
Confirm that:
- Inspiratory and expiratory sensors have not been reversed.
- Each sensor is fully inserted.
- Connections are secure.
- Seals are properly positioned.
- No sensor is visibly tilted, loose, or partially seated.
Expected outcome: Both sensors are installed correctly and form secure pneumatic connections.
Retry calibration. If it passes, complete the checkout and stop.
7. Check for Moisture in the Breathing System
Inspect accessible breathing-system areas for retained moisture, including:
- Sensor ports
- Breathing-system channels
- Water traps
- Reservoir bag connection
- Inspiratory and expiratory limbs
- Absorber assembly
- Condensation collection points
Drain or dry components according to the approved cleaning and reprocessing procedure.
Expected outcome: No visible moisture remains that could affect flow measurement.
Retry calibration after the system is fully dry. If calibration passes, complete the checkout and stop.
8. Check the Circuit and Test Connections
Verify that the breathing circuit is configured exactly as required by the on-screen calibration instructions.
Check for:
- Loose circuit connections
- Open circuit ports
- Missing plugs or caps
- Incorrect test-lung placement
- Damaged breathing hoses
- An open manual bag connection
- Loose absorber or canister components
- Incorrect APL valve or ventilation-selector position
Follow the prompts displayed by the specific WATO EX model and software version.
Expected outcome: The breathing system is sealed and configured correctly for calibration.
If correcting the setup resolves the failure, complete the checkout and stop.
9. Perform a System Leak Test
Run the machine’s approved system checkout or leak test.
If the leak test fails:
- Recheck the circuit connections.
- Inspect accessible seals and breathing-system components.
- Confirm the absorber assembly is properly installed.
- Check for cracked hoses or loose accessories.
- Correct the external leak before repeating flow-sensor calibration.
Expected outcome: The machine passes the leak test before flow calibration is attempted again.
If both the leak test and flow calibration pass, return the device to service only after completing all required performance checks.
10. Substitute Known-Good External Components
When approved components are available, substitute one item at a time:
- Known-good breathing circuit
- Known-good test lung
- Known-good compatible inspiratory flow sensor
- Known-good compatible expiratory flow sensor
Repeat calibration after each substitution.
Do not exchange the complete breathing-system block between machines unless authorized and prepared to perform the required calibration and verification.
Expected outcome: A defective external accessory or sensor is isolated without unnecessary internal repair.
If replacing a flow sensor resolves the failure, complete the full system checkout and verify volume delivery before returning the machine to service.
11. Power-Cycle and Repeat the Checkout
After confirming the machine is not connected to a patient:
- Shut down the system normally.
- Leave it powered off briefly.
- Restore verified AC power and gas supplies.
- Allow startup to finish completely.
- Repeat the system checkout and flow-sensor calibration.
Expected outcome: A temporary startup or calibration-state problem clears.
If the calibration passes, perform functional verification and document the final status.
If the Problem Persists
If calibration continues to fail after the sensors are clean, dry, correctly installed, the breathing system passes external inspection, and leaks have been ruled out, the problem is likely internal.
Possible internal causes include:
- Failed flow-sensor electronics
- Damaged sensor-interface connections
- Inspiratory-valve calibration problems
- Internal pneumatic leakage
- Breathing-system identification or matching problems
- Internal control or acquisition-board failure
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using the applicable Mindray service documentation and approved test equipment
Do not return the machine to clinical use based only on the disappearance of the message. Successful calibration, system checkout, leak testing, and verification of delivered and measured volumes are required.
Knowing when to stop and escalate is proper troubleshooting.
Clinical Use Tip
Never remove or calibrate flow sensors while the anesthesia machine is connected to an active patient. Transfer the patient first and troubleshoot only under controlled, nonclinical conditions.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Mindray WATO EX Series displayed a flow-sensor calibration failure during the pre-use checkout."
Cause
What was observed during troubleshooting.
Example:
"Inspection found moisture on the expiratory flow sensor and the sensor was not fully seated in the breathing-system assembly."
Resolution
What action was taken.
Example:
"Dried and correctly reinstalled the flow sensor, completed calibration, passed the system leak test and checkout, and verified stable volume measurements."
Helpful Details to Include (If Known)
- Complete displayed failure message
- Startup or checkout stage where failure occurred
- Inspiratory or expiratory sensor affected
- Moisture or contamination found
- Sensor orientation and seating checked
- Breathing-system block recently removed or exchanged
- Breathing circuit or test lung replaced
- Leak-test result
- Known-good sensors substituted
- Calibration result after corrective action
- Delivered and measured volume verification
- Final device status
Final Thought
Flow-sensor calibration failures should be approached by protecting the patient first, then checking moisture, sensor installation, breathing-system configuration, and external leaks. Escalate when calibration cannot be restored and document the complaint, confirmed cause, corrective action, and final verification clearly.
That is successful troubleshooting.