Mindray WATO EX Series

Flow Sensor Calibration Failure

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

What This Guide Helps With

Troubleshooting failed flow-sensor calibration caused by moisture, contamination, incorrect installation, breathing-system configuration, leaks, or unsuitable calibration conditions.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform flow-sensor calibration, remove breathing-system components, or disturb pneumatic connections while the WATO EX Series is supporting an active patient.

If the failure occurs during clinical use:

Expected outcome: Patient care no longer depends on equipment with uncertain flow or volume measurement.

Continue troubleshooting only after the machine has been safely removed from patient use. Mindray specifies that flow-sensor calibration must not be performed while the unit is connected to a patient.

2. Confirm the Exact Failure Condition

Record the complete displayed message and determine whether the issue involves:

Expected outcome: The failure is clearly documented and can be reproduced safely.

If the calibration completes successfully and the system checkout passes, stop troubleshooting.

3. Allow the Machine to Complete Startup

Verify that:

Do not operate pneumatic controls, adjust gas flows, or disturb the breathing system while calibration is running.

Expected outcome: The machine is stable and ready to perform an uninterrupted calibration.

If calibration now passes, document the result and stop.

4. Inspect the Breathing System Externally

Confirm that the breathing-system assembly is:

Some WATO EX configurations are calibrated to their associated breathing-system block. Replacing or exchanging that assembly may require flow-sensor recalibration.

Expected outcome: The correct breathing system is properly installed without visible mechanical misalignment.

If reseating the assembly resolves the calibration failure, complete the system checkout and stop.

5. Inspect the Flow Sensors

Following facility procedures and the applicable Mindray instructions, remove and inspect the accessible inspiratory and expiratory flow sensors.

Check for:

Do not insert tools into the sensor, scrape sensing elements, or use unapproved cleaning methods.

Expected outcome: Both sensors are clean, dry, undamaged, and suitable for reinstallation.

If contamination or moisture is found, clean and dry the components only according to the approved procedure before continuing.

6. Verify Correct Sensor Installation

Reinstall each sensor in its designated location and orientation.

Confirm that:

Expected outcome: Both sensors are installed correctly and form secure pneumatic connections.

Retry calibration. If it passes, complete the checkout and stop.

7. Check for Moisture in the Breathing System

Inspect accessible breathing-system areas for retained moisture, including:

Drain or dry components according to the approved cleaning and reprocessing procedure.

Expected outcome: No visible moisture remains that could affect flow measurement.

Retry calibration after the system is fully dry. If calibration passes, complete the checkout and stop.

8. Check the Circuit and Test Connections

Verify that the breathing circuit is configured exactly as required by the on-screen calibration instructions.

Check for:

Follow the prompts displayed by the specific WATO EX model and software version.

Expected outcome: The breathing system is sealed and configured correctly for calibration.

If correcting the setup resolves the failure, complete the checkout and stop.

9. Perform a System Leak Test

Run the machine’s approved system checkout or leak test.

If the leak test fails:

Expected outcome: The machine passes the leak test before flow calibration is attempted again.

If both the leak test and flow calibration pass, return the device to service only after completing all required performance checks.

10. Substitute Known-Good External Components

When approved components are available, substitute one item at a time:

Repeat calibration after each substitution.

Do not exchange the complete breathing-system block between machines unless authorized and prepared to perform the required calibration and verification.

Expected outcome: A defective external accessory or sensor is isolated without unnecessary internal repair.

If replacing a flow sensor resolves the failure, complete the full system checkout and verify volume delivery before returning the machine to service.

11. Power-Cycle and Repeat the Checkout

After confirming the machine is not connected to a patient:

Expected outcome: A temporary startup or calibration-state problem clears.

If the calibration passes, perform functional verification and document the final status.

If the Problem Persists

If calibration continues to fail after the sensors are clean, dry, correctly installed, the breathing system passes external inspection, and leaks have been ruled out, the problem is likely internal.

Possible internal causes include:

The device should be:

Do not return the machine to clinical use based only on the disappearance of the message. Successful calibration, system checkout, leak testing, and verification of delivered and measured volumes are required.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never remove or calibrate flow sensors while the anesthesia machine is connected to an active patient. Transfer the patient first and troubleshoot only under controlled, nonclinical conditions.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Mindray WATO EX Series displayed a flow-sensor calibration failure during the pre-use checkout."

Cause

What was observed during troubleshooting.

Example:
"Inspection found moisture on the expiratory flow sensor and the sensor was not fully seated in the breathing-system assembly."

Resolution

What action was taken.

Example:
"Dried and correctly reinstalled the flow sensor, completed calibration, passed the system leak test and checkout, and verified stable volume measurements."

Helpful Details to Include (If Known)

Final Thought

Flow-sensor calibration failures should be approached by protecting the patient first, then checking moisture, sensor installation, breathing-system configuration, and external leaks. Escalate when calibration cannot be restored and document the complaint, confirmed cause, corrective action, and final verification clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide