Mindray WATO EX Series

Fresh Gas Flow Control Failure

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

What This Guide Helps With

Troubleshooting unavailable, unstable, or incorrect fresh-gas flow caused by gas supplies, flow controls, settings, startup conditions, or external connections.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended fresh-gas flow troubleshooting while the WATO EX Series anesthesia machine is supporting an active patient.

If fresh-gas delivery becomes unavailable, unstable, or inconsistent during clinical use:

Expected outcome: Patient care no longer depends on equipment with uncertain fresh-gas delivery.

Continue troubleshooting only after the machine has been safely removed from active patient use.

2. Confirm the Exact Failure Condition

Record the complete displayed alarm, advisory, or startup-test message.

Determine whether:

Expected outcome: The affected gas, control, and operating condition are clearly identified.

If the reported condition cannot be reproduced, review alarm logs and obtain details from the clinical user before returning the unit to service.

3. Verify Medical Gas Availability

Confirm that the required pipeline hoses are:

Check the machine’s displayed supply-pressure indicators for oxygen, air, and nitrous oxide, as equipped.

If cylinders are installed:

The WATO EX Series requires an adequate drive-gas and medical-gas supply for normal operation; published WATO EX-65 specifications list a ventilator driving-pressure range of 280–600 kPa.

Expected outcome: Required gases show normal supply pressure without supply alarms.

If the issue is corrected after restoring the gas supply, complete the operational checkout and stop troubleshooting.

4. Isolate the Pipeline Supply

Connect the machine to a different verified wall outlet using an approved gas hose, when available.

Alternatively, test temporarily from a verified backup cylinder in accordance with facility policy.

If flow operates normally from the alternate source, investigate the original pipeline outlet, hose, connector, or facility gas supply.

If the failure remains with both pipeline and cylinder sources, continue troubleshooting the anesthesia machine.

Expected outcome: The facility supply is either identified as the cause or ruled out.

5. Inspect the Flow-Control Interface

Depending on the WATO EX model, fresh-gas flow may be adjusted using mechanical flow controls or an electronic digital gas mixer. Mindray describes the WATO EX-65 Pro as using a digital gas mixer with total-flow and direct-flow control modes.

Check that:

Do not force a stiff flow-control knob or repeatedly operate an unresponsive electronic control.

Expected outcome: Controls move normally and corresponding settings change on the display.

If a damaged, loose, or binding control is found, remove the device from service for repair.

6. Verify Fresh-Gas Settings

Confirm that the selected settings are clinically reasonable and not unintentionally set to minimum or zero flow.

Check:

The optimizer is guidance for fresh-gas efficiency; it does not by itself prove that the gas-delivery system is functioning correctly. Mindray documentation describes the optimizer as indicating whether fresh-gas flow is insufficient, appropriate, or excessive.

Expected outcome: Selected flow and gas-mixture values are appropriate and accepted by the machine.

If correcting an unintended setting restores normal delivery, perform a complete checkout and stop troubleshooting.

7. Check for Hypoxic-Guard or Gas-Supply Limitations

Verify whether the machine is intentionally limiting nitrous oxide or total gas flow because of:

Restore a verified oxygen supply before evaluating other gas-flow controls.

Do not attempt to bypass oxygen-ratio protection or other safety functions.

Expected outcome: Gas-ratio safety functions permit the selected mixture after adequate oxygen supply is restored.

8. Inspect the Common Gas Path and External Connections

Check externally accessible components for incorrect setup or obstruction:

Confirm that gas is being directed to the intended outlet and not unintentionally diverted to an auxiliary outlet.

Expected outcome: The fresh-gas pathway is correctly assembled, unobstructed, and directed to the intended breathing system.

If correcting a connection restores proper flow, complete the machine checkout and stop troubleshooting.

9. Inspect the Vaporizer Installation

Confirm that each vaporizer is:

Turn all vaporizers off and determine whether normal carrier-gas flow can be established.

Do not remove or install a vaporizer while gas is actively flowing unless permitted by the manufacturer’s instructions.

Expected outcome: Normal fresh-gas flow is available with vaporizers off and correctly mounted.

If flow failure occurs only when a specific vaporizer is activated, remove that vaporizer from use and evaluate it separately.

10. Perform a Controlled Restart

After recording all displayed messages:

Do not repeatedly power-cycle the machine to clear a persistent fault.

Expected outcome: The system completes startup without a fresh-gas control or internal communication fault.

If normal operation returns, continue with a complete pre-use checkout before considering the device ready for service.

11. Run the Approved System Checkout

Perform the operator-level pre-use checkout or leak/compliance test specified for the exact WATO EX model.

Before testing:

Mindray service documentation indicates that detected fresh-gas flow can prevent completion of certain leak tests and may prompt the operator to zero all flowmeters.

Expected outcome: The machine passes its startup, leak, compliance, and gas-delivery checks without recurring flow-control messages.

Do not return the device to clinical use if the checkout fails or must be overridden.

12. Verify Delivered Flow With Approved Test Equipment

Using an anesthesia-machine analyzer or calibrated flow-measurement device:

Use the acceptance limits and test method specified in the applicable Mindray service documentation and facility procedure.

Expected outcome: Measured flow and oxygen concentration remain stable and within the manufacturer’s permitted tolerances.

If displayed settings do not agree with independently measured output, remove the machine from service.

If the Problem Persists

If verified gas supplies, hoses, settings, controls, vaporizers, external connections, and the breathing-system setup have been ruled out, the failure is likely internal.

Possible internal causes include:

The anesthesia machine should be:

Do not perform internal pneumatic adjustments, valve replacement, calibration, or board-level repair without the appropriate Mindray service documentation, training, and test equipment.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Never troubleshoot uncertain fresh-gas delivery on an active patient. Transfer the patient to another verified anesthesia machine before changing gas supplies, vaporizers, flow controls, or system configuration.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Mindray WATO EX Series machine would not increase total fresh-gas flow above zero."

Cause

What was observed during troubleshooting.

Example:
"Oxygen pipeline pressure was absent because the oxygen hose was not fully seated in the wall outlet, causing the machine to restrict gas delivery."

Resolution

What action was taken.

Example:
"Reseated the oxygen pipeline hose, confirmed normal supply pressure, verified fresh-gas flow and oxygen concentration with an analyzer, completed the system checkout, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Fresh-gas flow failures can directly affect oxygen and anesthetic delivery. Protect the patient first, verify external supplies and configuration logically, measure actual output independently, and escalate persistent faults with complete CCR documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide