Mindray WATO EX Series

High Airway Pressure Alarm

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

What This Guide Helps With

Troubleshooting elevated airway pressure caused by circuit restrictions, patient-side accessories, valve problems, ventilator settings, or pressure-monitoring faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the WATO EX Series is supporting an active patient.

If a high airway pressure alarm occurs during clinical use:

Expected outcome: Patient ventilation remains safely supported without depending on equipment with unresolved pressure behavior.

Continue troubleshooting only after the machine has been removed from active patient use.

2. Confirm the Exact Alarm Condition

Record:

The WATO EX Series monitors airway pressure and provides an adjustable high-pressure alarm limit; exact ranges and available ventilation settings vary by EX model.

Expected outcome: The alarm is confirmed as a reproducible airway-pressure condition rather than an isolated clinical event.

If the alarm cannot be reproduced and the system passes checkout, document the reported conditions and monitor the device according to facility policy.

3. Check the High-Pressure Alarm Limit

Verify that the airway high-pressure alarm limit is:

Do not raise the alarm limit simply to silence a recurring alarm.

Expected outcome: The alarm limit is appropriately configured and not producing nuisance alarms during otherwise normal test ventilation.

If correcting the setting resolves the issue, complete a full checkout and stop.

4. Inspect the Patient Breathing Circuit

Examine the complete circuit from the machine to the patient connection for:

Replace questionable components with known-good, approved accessories.

Expected outcome: The breathing pathway is open, correctly assembled, and free of visible restrictions.

If replacing a restricted circuit component resolves the alarm, complete checkout and stop.

5. Remove Patient-Side Accessories

Disconnect nonessential patient-side accessories such as:

Install a simple known-good test circuit and test lung.

Expected outcome: The machine ventilates the test lung without excessive peak pressure.

If the alarm clears, reconnect accessories individually to identify the restricted or incompatible component.

6. Check the Test Lung and Patient Connection

When bench testing:

Expected outcome: The test load expands normally and the pressure waveform rises and falls predictably.

If a defective test lung or blocked connector caused the alarm, replace it and stop after successful checkout.

7. Inspect the Inspiratory and Expiratory Paths

With the machine removed from service, inspect accessible breathing-system components for:

Do not perform internal disassembly beyond procedures authorized by the facility and manufacturer.

Expected outcome: Accessible valves and breathing-system components are clean, dry, correctly installed, and move freely.

If correcting an installation problem resolves the alarm, perform the required checkout and leak test before returning the device to service.

8. Verify the APL Valve and Ventilation Mode

Confirm whether the alarm occurs in:

In manual mode, verify that the adjustable pressure-limiting valve:

Expected outcome: Pressure can be controlled and released normally in manual mode.

A problem limited to manual mode may indicate an APL valve or manual breathing-path issue. A problem limited to mechanical ventilation may indicate a ventilator, expiratory-valve, or control problem.

9. Review Ventilator Settings

Verify that the selected settings are appropriate for the connected test lung:

Excessive tidal volume, inspiratory pressure, PEEP, or inspiratory time can produce elevated measured airway pressure.

Expected outcome: Test ventilation operates normally with controlled, appropriate settings.

If correcting an unintended setting resolves the alarm, complete checkout and stop.

10. Check for Incomplete Exhalation

Observe whether airway pressure returns to the selected PEEP level before the next breath.

Check for:

Expected outcome: The test lung fully exhales and pressure returns to baseline between breaths.

If correcting the setting or expiratory restriction resolves the alarm, perform a complete operational checkout and stop.

11. Check the Scavenging Connection

Inspect the waste-gas scavenging system for:

Temporarily test using the facility-approved configuration described in the applicable service documentation. Do not vent anesthetic gas into an occupied workspace.

Expected outcome: Scavenging operates without creating abnormal back pressure in the breathing system.

If the external scavenging system caused the alarm, correct the external condition and complete checkout.

12. Perform the System Checkout

After correcting external issues:

Expected outcome: The machine passes checkout and delivers stable ventilation without unexpected high-pressure alarms.

If checkout passes and the alarm does not recur, document the repair and return the device according to facility policy.

If the Problem Persists

If the alarm continues with a known-good circuit, test lung, accessories, appropriate settings, unobstructed scavenging system, and correctly installed breathing-system components, common external causes have been ruled out.

The issue may involve:

The device should be:

Do not return the machine to clinical use based only on the alarm clearing after repeated restarts. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a recurring airway-pressure alarm on an active patient beyond immediate safety checks. Move the patient to another verified ventilation method before testing circuits, valves, settings, or internal functions.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported repeated high airway pressure alarms during mechanical ventilation on the Mindray WATO EX Series."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the expiratory breathing hose kinked behind the absorber assembly, restricting exhalation and causing pressure to remain elevated."

Resolution

What action was taken.

Example:
"Repositioned and replaced the affected hose, completed the system checkout, and verified stable pressure and normal alarm operation using a test lung."

Helpful Details to Include (If Known)

Final Thought

High airway pressure can result from a simple kinked hose or a more serious ventilation-control fault. Protect the patient first, isolate external restrictions logically, verify operation through a complete checkout, and escalate any unresolved pressure abnormality. Accurate CCR documentation preserves the findings and supports safe follow-up.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide