Mindray WATO EX Series

O2 Sensor Calibration or FiO2 Measurement Error

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

What This Guide Helps With

Troubleshooting failed O2 sensor calibration, missing or inaccurate FiO2 readings, unstable measurements, sensor connection problems, moisture, or gas-supply conditions.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not calibrate, disconnect, remove, or replace oxygen-measurement components while the WATO EX Series is supporting an active patient.

If FiO2 is unavailable, unstable, or inconsistent with the selected gas mixture:

Expected outcome: Patient care no longer depends on equipment with questionable oxygen measurement.

Continue troubleshooting only after the machine has been safely removed from active patient use.

2. Confirm the Exact Failure Condition

Record the complete displayed alarm, advisory, or calibration message.

Determine whether the issue involves:

Expected outcome: The affected measurement source and failure condition are clearly identified.

3. Confirm Which Oxygen Measurement Source Is Active

Verify whether the machine is using:

Review the displayed measurement labels and configuration without changing clinical settings unnecessarily.

Some WATO EX configurations may use an oxygen sensor, a gas module, or both. Calibration availability can depend on which monitoring source is enabled.

Expected outcome: Clinical Engineering knows which sensor or module is producing the questionable FiO2 value.

4. Inspect the O2 Sensor and External Connection

With the machine powered down or placed in the appropriate safe standby condition:

A disconnected or poorly connected sensor can cause calibration to stop immediately or generate a sensor-disconnected indication.

Expected outcome: The sensor is correctly seated, connected, and recognized.

If this resolves the issue, complete the required checkout and stop troubleshooting.

5. Inspect for Moisture or Contamination

Remove the O2 sensor only when permitted by the applicable operator instructions and facility procedure.

Inspect the sensor port and exposed sensor surfaces for:

If moisture is present:

Moisture accumulation can interfere with oxygen-sensor performance and may require drying before recalibration.

Expected outcome: The sensor and sensor port are clean, dry, and undamaged.

6. Verify the Breathing-System Setup

Inspect the breathing system for external setup conditions that could prevent the sensor from receiving the expected calibration gas:

Expected outcome: Calibration gas can reach the oxygen sensor without a significant leak or incorrect gas path.

7. Verify the Oxygen Supply

Confirm that a reliable oxygen source is connected.

Check:

Insufficient oxygen supply pressure can cause 100% oxygen calibration to fail.

Expected outcome: The anesthesia machine has a stable and adequate oxygen supply.

8. Allow the Machine and Sensor to Stabilize

If the machine was recently:

Allow sufficient time for the system and sensor to reach a stable operating condition.

Observe whether the FiO2 value stops drifting before attempting calibration again.

Expected outcome: The sensor output becomes stable enough for calibration.

9. Perform the 21% O2 Calibration

Place the machine in the required nonclinical condition and expose the oxygen sensor to clean room air as directed by the applicable WATO EX operating instructions.

Confirm:

Initiate the user-accessible 21% O2 calibration from the appropriate calibration menu.

Do not interrupt the calibration or change the breathing-system setup while it is running.

Expected outcome: The machine accepts room air as approximately 21% oxygen and reports successful calibration.

If the calibration is successful, reinstall or return the sensor to its normal operating position as required, complete system checkout, and stop troubleshooting. Mindray documentation indicates that successful 21% calibration precedes sensor reinstallation on applicable configurations.

10. Perform 100% O2 Calibration Only When Required

Perform 100% calibration only when:

Mindray documentation for applicable WATO EX-55/65 configurations states that 100% calibration must be performed in standby and only after successful 21% calibration.

Allow the system to flush fully with oxygen and wait for the reading to stabilize before starting calibration.

Expected outcome: The machine accepts the verified high-oxygen reference and completes calibration.

Do not repeatedly force calibration when the displayed oxygen value remains unstable.

11. Compare the Reading With an Independent Analyzer

After successful calibration:

Consider the specified accuracy and response time of both analyzers.

Expected outcome: The internal FiO2 measurement agrees with the independent analyzer within the applicable manufacturer and facility tolerance.

If the internal reading remains significantly inaccurate, stop testing and escalate the device.

12. Evaluate the Sensor’s Condition and Service Life

If calibration repeatedly fails despite correct setup, dry conditions, stable gas supply, and secure connections, assess whether the electrochemical O2 sensor may be depleted or defective.

Look for:

Replace the O2 sensor only with the correct approved part and according to facility policy.

After replacement:

Expected outcome: A known-good sensor calibrates successfully and provides accurate, stable FiO2 measurements.

If a known-good sensor does not resolve the problem, internal evaluation is required.

13. Complete the Full System Checkout

Before returning the WATO EX Series to service:

Expected outcome: Oxygen measurement and alarm functions pass all required tests.

Do not return the machine to clinical service based only on a successful calibration message.

If the Problem Persists

If the O2 sensor remains unrecognized, calibration repeatedly fails, or the displayed FiO2 remains inaccurate after the external checks, common setup and accessory causes have been ruled out.

The fault may involve:

The device should be:

Do not perform internal board-level repair or unauthorized calibration adjustments during routine troubleshooting.

Knowing when to stop and escalate is proper Clinical Engineering troubleshooting.

Clinical Use Tip

Do not troubleshoot the WATO EX Series while it is supporting an active patient. Move the patient to a verified backup anesthesia machine or approved ventilation system first to maintain therapy continuity.

Never use pulse oximetry alone to confirm the oxygen concentration delivered by an anesthesia machine. Move the patient to verified equipment and use an independent oxygen analyzer whenever FiO2 accuracy is uncertain.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the WATO EX Series displayed an O2 sensor calibration failure and an FiO2 value that did not match the selected oxygen mixture."

Cause

What was observed during troubleshooting.

Example:
"Inspection found condensation around the breathing-system O2 sensor and an unstable sensor reading that prevented successful 21% calibration."

Resolution

What action was taken.

Example:
"Removed the machine from service, dried and reseated the O2 sensor, completed successful calibration, verified FiO2 against a calibrated external analyzer, and passed the full system checkout."

Helpful Details to Include (If Known)

Final Thought

Oxygen-measurement faults require a patient-safety-first response and independent verification. Check the sensor, moisture, connections, breathing-system setup, and gas supply before suspecting internal failure. Escalate when calibration or accuracy cannot be verified, and document the complaint, identified cause, corrective action, test results, and final equipment status.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide