On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting failed O2 sensor calibration, missing or inaccurate FiO2 readings, unstable measurements, sensor connection problems, moisture, or gas-supply conditions.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not calibrate, disconnect, remove, or replace oxygen-measurement components while the WATO EX Series is supporting an active patient.
If FiO2 is unavailable, unstable, or inconsistent with the selected gas mixture:
- Notify the anesthesia provider immediately.
- Confirm oxygen delivery using an approved independent oxygen analyzer.
- Maintain independent patient monitoring, including pulse oximetry and respiratory-gas monitoring.
- Move the patient to another verified anesthesia machine or approved ventilation system if oxygen delivery cannot be confirmed.
- Follow the facility’s anesthesia-equipment failure procedure.
Expected outcome: Patient care no longer depends on equipment with questionable oxygen measurement.
Continue troubleshooting only after the machine has been safely removed from active patient use.
2. Confirm the Exact Failure Condition
Record the complete displayed alarm, advisory, or calibration message.
Determine whether the issue involves:
- O2 sensor calibration failure
- O2 sensor disconnected message
- FiO2 measurement unavailable
- FiO2 reading fixed, drifting, or unstable
- Displayed FiO2 not matching the selected gas mixture
- Failure during 21% calibration
- Failure during 100% calibration
- A discrepancy between the internal O2 sensor and an external gas analyzer
Expected outcome: The affected measurement source and failure condition are clearly identified.
3. Confirm Which Oxygen Measurement Source Is Active
Verify whether the machine is using:
- The breathing-system oxygen sensor
- An installed respiratory-gas module
- Another configured oxygen-measurement source
Review the displayed measurement labels and configuration without changing clinical settings unnecessarily.
Some WATO EX configurations may use an oxygen sensor, a gas module, or both. Calibration availability can depend on which monitoring source is enabled.
Expected outcome: Clinical Engineering knows which sensor or module is producing the questionable FiO2 value.
4. Inspect the O2 Sensor and External Connection
With the machine powered down or placed in the appropriate safe standby condition:
- Confirm the O2 sensor is fully installed in its designated breathing-system port.
- Inspect the sensor cable and connector for looseness, bent contacts, contamination, or damage.
- Confirm the cable is not pinched, stretched, or routed where it can be pulled.
- Disconnect and firmly reconnect accessible sensor connections when permitted by facility procedure.
- Check for an “O2 Sensor Disconnected” message after reconnection.
A disconnected or poorly connected sensor can cause calibration to stop immediately or generate a sensor-disconnected indication.
Expected outcome: The sensor is correctly seated, connected, and recognized.
If this resolves the issue, complete the required checkout and stop troubleshooting.
5. Inspect for Moisture or Contamination
Remove the O2 sensor only when permitted by the applicable operator instructions and facility procedure.
Inspect the sensor port and exposed sensor surfaces for:
- Condensation
- Water droplets
- Cleaning solution residue
- Breathing-circuit contamination
- Cracks or physical damage
If moisture is present:
- Remove visible moisture from the sensor area.
- Allow the sensor to air-dry completely.
- Do not apply compressed air, heat, solvents, or unapproved cleaning chemicals.
- Inspect the breathing system for the source of the moisture.
Moisture accumulation can interfere with oxygen-sensor performance and may require drying before recalibration.
Expected outcome: The sensor and sensor port are clean, dry, and undamaged.
6. Verify the Breathing-System Setup
Inspect the breathing system for external setup conditions that could prevent the sensor from receiving the expected calibration gas:
- Breathing system fully assembled
- Absorber canister correctly installed
- Inspiratory and expiratory components properly seated
- Circuit connections secure
- Manual bag or test lung connected as required
- APL valve and bag/vent switch positioned according to the calibration instructions
- No obvious breathing-circuit leak
- No open auxiliary gas outlet diverting gas away from the breathing system
Expected outcome: Calibration gas can reach the oxygen sensor without a significant leak or incorrect gas path.
7. Verify the Oxygen Supply
Confirm that a reliable oxygen source is connected.
Check:
- Pipeline O2 hose connected to the correct wall outlet
- Pipeline pressure within the facility’s acceptable range
- Cylinder supply available when used as the test source
- Cylinder valve open and pressure adequate
- No O2 supply-pressure alarm
- No kinked or damaged supply hose
Insufficient oxygen supply pressure can cause 100% oxygen calibration to fail.
Expected outcome: The anesthesia machine has a stable and adequate oxygen supply.
8. Allow the Machine and Sensor to Stabilize
If the machine was recently:
- Powered on
- Moved between areas with different temperatures
- Cleaned
- Exposed to condensation
- Stored without use
- Fitted with a replacement sensor
Allow sufficient time for the system and sensor to reach a stable operating condition.
Observe whether the FiO2 value stops drifting before attempting calibration again.
Expected outcome: The sensor output becomes stable enough for calibration.
9. Perform the 21% O2 Calibration
Place the machine in the required nonclinical condition and expose the oxygen sensor to clean room air as directed by the applicable WATO EX operating instructions.
Confirm:
- No patient is connected.
- The sensor is exposed to uncontaminated room air.
- Oxygen or nitrous oxide is not flowing into the sensor area.
- Residual anesthetic gas has been cleared.
- The displayed value is stable before calibration begins.
Initiate the user-accessible 21% O2 calibration from the appropriate calibration menu.
Do not interrupt the calibration or change the breathing-system setup while it is running.
Expected outcome: The machine accepts room air as approximately 21% oxygen and reports successful calibration.
If the calibration is successful, reinstall or return the sensor to its normal operating position as required, complete system checkout, and stop troubleshooting. Mindray documentation indicates that successful 21% calibration precedes sensor reinstallation on applicable configurations.
10. Perform 100% O2 Calibration Only When Required
Perform 100% calibration only when:
- It is required by the applicable model instructions or facility procedure.
- The 21% calibration has completed successfully.
- The machine is in standby or the required nonclinical mode.
- A verified 100% oxygen supply is available.
- Clinical Engineering is authorized to access the applicable calibration menu.
Mindray documentation for applicable WATO EX-55/65 configurations states that 100% calibration must be performed in standby and only after successful 21% calibration.
Allow the system to flush fully with oxygen and wait for the reading to stabilize before starting calibration.
Expected outcome: The machine accepts the verified high-oxygen reference and completes calibration.
Do not repeatedly force calibration when the displayed oxygen value remains unstable.
11. Compare the Reading With an Independent Analyzer
After successful calibration:
- Connect a calibrated external oxygen analyzer at an appropriate breathing-system test point.
- Deliver test gas using approved functional-test settings.
- Compare the WATO EX FiO2 reading with the independent analyzer.
- Test at room-air concentration and at a higher oxygen concentration when permitted.
- Allow both measurements to stabilize before comparing them.
Consider the specified accuracy and response time of both analyzers.
Expected outcome: The internal FiO2 measurement agrees with the independent analyzer within the applicable manufacturer and facility tolerance.
If the internal reading remains significantly inaccurate, stop testing and escalate the device.
12. Evaluate the Sensor’s Condition and Service Life
If calibration repeatedly fails despite correct setup, dry conditions, stable gas supply, and secure connections, assess whether the electrochemical O2 sensor may be depleted or defective.
Look for:
- Slow response to changes in oxygen concentration
- Output that remains fixed
- Excessive drift
- Repeated calibration failure
- Failure at both 21% and 100%
- Physical damage or leakage
- Sensor age beyond the facility’s expected replacement interval
Replace the O2 sensor only with the correct approved part and according to facility policy.
After replacement:
- Allow the sensor to stabilize.
- Perform the required calibration.
- Complete the full anesthesia-machine checkout.
- Verify performance with an independent oxygen analyzer.
Expected outcome: A known-good sensor calibrates successfully and provides accurate, stable FiO2 measurements.
If a known-good sensor does not resolve the problem, internal evaluation is required.
13. Complete the Full System Checkout
Before returning the WATO EX Series to service:
- Complete the manufacturer-required preoperative or system checkout.
- Confirm the FiO2 value is present and stable.
- Verify low- and high-O2 alarm operation when included in the approved test procedure.
- Confirm the reading responds appropriately when oxygen concentration changes.
- Compare the reading with an independent calibrated analyzer.
- Confirm no sensor-disconnected or calibration messages remain.
- Inspect the breathing system for leaks.
- Document the final device status.
Expected outcome: Oxygen measurement and alarm functions pass all required tests.
Do not return the machine to clinical service based only on a successful calibration message.
If the Problem Persists
If the O2 sensor remains unrecognized, calibration repeatedly fails, or the displayed FiO2 remains inaccurate after the external checks, common setup and accessory causes have been ruled out.
The fault may involve:
- A depleted or defective O2 sensor
- Damaged sensor wiring or connector
- Sensor-interface circuitry
- Internal pneumatic delivery instability
- Gas-module failure
- Measurement or control electronics
- Calibration data or software configuration
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for qualified repair or bench evaluation
- Evaluated using the correct Mindray service documentation and test equipment
Do not perform internal board-level repair or unauthorized calibration adjustments during routine troubleshooting.
Knowing when to stop and escalate is proper Clinical Engineering troubleshooting.
Clinical Use Tip
Do not troubleshoot the WATO EX Series while it is supporting an active patient. Move the patient to a verified backup anesthesia machine or approved ventilation system first to maintain therapy continuity.
Never use pulse oximetry alone to confirm the oxygen concentration delivered by an anesthesia machine. Move the patient to verified equipment and use an independent oxygen analyzer whenever FiO2 accuracy is uncertain.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Anesthesia staff reported that the WATO EX Series displayed an O2 sensor calibration failure and an FiO2 value that did not match the selected oxygen mixture."
Cause
What was observed during troubleshooting.
Example:
"Inspection found condensation around the breathing-system O2 sensor and an unstable sensor reading that prevented successful 21% calibration."
Resolution
What action was taken.
Example:
"Removed the machine from service, dried and reseated the O2 sensor, completed successful calibration, verified FiO2 against a calibrated external analyzer, and passed the full system checkout."
Helpful Details to Include (If Known)
- Complete alarm or calibration message recorded
- Oxygen-measurement source identified
- O2 sensor connection inspected
- Sensor port checked for moisture
- Breathing-system setup verified
- Pipeline O2 pressure checked
- Cylinder pressure checked
- 21% calibration attempted
- 100% calibration attempted, when applicable
- External oxygen analyzer used
- Internal and external readings recorded
- Replacement sensor tested
- Alarm behavior documented
- Accessories swapped or tested
- Power behavior documented
- Environmental factors documented
- Indicator lights documented
- Unusual sounds noted
- Unusual heat or smell noted
- Full system checkout completed
- Final device status documented
Final Thought
Oxygen-measurement faults require a patient-safety-first response and independent verification. Check the sensor, moisture, connections, breathing-system setup, and gas supply before suspecting internal failure. Escalate when calibration or accuracy cannot be verified, and document the complaint, identified cause, corrective action, test results, and final equipment status.
That is successful troubleshooting.