Mindray WATO EX Series

PEEP High or Elevated Expiratory Pressure

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

What This Guide Helps With

Troubleshooting elevated measured PEEP or incomplete exhalation caused by settings, circuit restrictions, breathing-system components, scavenging, or external accessory problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the WATO EX Series anesthesia machine is supporting an active patient.

If elevated PEEP or expiratory pressure occurs during a procedure:

Expected outcome: Patient care no longer depends on equipment with unresolved expiratory-pressure performance.

Continue troubleshooting only after the machine has been removed from active patient use.

2. Confirm the Reported Pressure Condition

Record:

The WATO EX Series uses electronically controlled PEEP on applicable configurations, with model-dependent settings and monitoring capabilities.

Expected outcome: The exact difference between selected and measured PEEP is identified.

If the measured PEEP matches the selected setting and the device passes checkout, confirm the intended clinical setting with the anesthesia provider before proceeding.

3. Verify the PEEP and Ventilation Settings

Confirm that:

High respiratory rates or insufficient expiratory time can prevent the test lung from fully emptying and create apparent auto-PEEP.

Expected outcome: Pressure returns to the selected baseline between breaths.

If correcting an unintended setting resolves the issue, complete a full system checkout and stop troubleshooting.

4. Inspect the Patient Breathing Circuit

Remove and inspect the complete breathing circuit for:

Replace questionable components with known-good compatible accessories.

Expected outcome: The circuit is open, correctly assembled, and free of restrictions.

If replacing the circuit resolves the elevated pressure, complete checkout and return the machine to service only after satisfactory testing.

5. Remove Unnecessary External Accessories

Temporarily remove or replace, as appropriate:

Use a manufacturer-compatible test circuit and test lung.

Expected outcome: Measured PEEP returns to the selected value when the restrictive accessory is removed.

Do not return a visibly contaminated, wet, damaged, or restricted accessory to service.

6. Check the Expiratory Limb and Expiratory Port

Inspect the expiratory limb from the patient connection back to the breathing system.

Confirm that:

Do not insert tools into ports or attempt internal valve disassembly.

Expected outcome: Exhaled gas can move freely through the expiratory pathway.

If reseating an externally accessible component resolves the issue, perform the required checkout before returning the device to service.

7. Inspect the Breathing System Assembly

With the device removed from patient use, visually inspect accessible components for:

Compare the assembly with another verified WATO EX unit when available.

Expected outcome: The breathing system is completely assembled with no visible obstruction or misalignment.

8. Evaluate the Scavenging Connection

Inspect the anesthetic gas scavenging system for:

Test the facility scavenging connection independently when permitted by facility procedure.

Expected outcome: The scavenging system removes waste gas without applying abnormal pressure or suction to the breathing system.

If the fault follows the wall connection or scavenging hose, correct the facility-side problem and repeat checkout.

9. Compare Mechanical and Electronic Pressure Information

Using an approved breathing-circuit analyzer or pressure meter, compare independently measured pressure with the machine’s displayed PEEP.

Confirm that:

Expected outcome: Independent and displayed pressure measurements agree within the applicable acceptance criteria.

If independent pressure is normal but the displayed PEEP remains elevated, suspect a pressure-measurement or calibration problem.

If both measurements are elevated, suspect an actual expiratory restriction or PEEP-control problem.

10. Run the Normal Preuse or System Checkout

Restore the machine to its standard approved test configuration and run the complete normal checkout.

Do not bypass failed tests or repeatedly clear alarms without identifying the cause.

Record:

Expected outcome: The machine passes checkout and maintains the selected PEEP without abnormal residual pressure.

If the machine passes repeatedly and measured pressures are within specifications, return it to service according to facility policy.

If the Problem Persists

If elevated PEEP remains after settings, breathing circuits, accessories, flow sensors, breathing-system assembly, scavenging, and external measurement have been checked, common external causes have been ruled out.

The problem may involve the internal expiratory valve, electronically controlled PEEP system, pressure-sensing pathway, flow measurement, pneumatic control, or ventilator electronics.

The device should be:

Do not continue clinical use based only on the machine completing ventilation when expiratory pressure remains abnormal. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot elevated PEEP on an active patient beyond the immediate checks needed to protect the patient. Transfer ventilation first, then evaluate the machine using a test lung and calibrated analyzer.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the WATO EX Series displayed PEEP above the selected value and airway pressure did not return fully to baseline."

Cause

What was observed during troubleshooting.

Example:
"Inspection found water accumulation and partial obstruction in the expiratory breathing-circuit limb, causing increased resistance during exhalation."

Resolution

What action was taken.

Example:
"Replaced the affected circuit, inspected the expiratory pathway, completed system checkout, and verified selected and measured PEEP using a calibrated breathing-circuit analyzer."

Helpful Details to Include (If Known)

Final Thought

Elevated expiratory pressure can result from a simple wet filter or circuit obstruction, but it can also indicate a failed pressure-control component. Protect the patient first, verify the full expiratory pathway logically, compare pressures independently, and escalate when normal operation cannot be proven. Accurate CCR documentation supports safe repair and prevents repeated troubleshooting.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide