Mindray WATO EX Series

Pre-Use Check or Leak Test Failure

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

What This Guide Helps With

Troubleshooting failed pre-use or breathing-system leak tests caused by circuit setup, loose connections, open valves, accessories, seals, or test preparation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a failed pre-use check or leak test while the WATO EX Series is supporting an active patient.

If the failure is discovered during clinical use:

Expected outcome: Patient care is no longer dependent on an anesthesia machine that has not successfully completed its required checks.

Continue troubleshooting only after the machine has been removed from active patient use.

2. Record the Exact Failed Check

Review the displayed test result and determine whether the failure involves:

Record the complete message, test stage, and any displayed corrective guidance. Mindray anesthesia systems identify failed automatic or manual circuit leak tests through their system-check functions.

Expected outcome: The failed portion of the checkout is identified rather than treating every pre-use failure as the same problem.

3. Confirm Adequate Gas and Electrical Supplies

Verify:

Correct any external supply issue and repeat the check.

Expected outcome: The machine has stable power and adequate gas supplies throughout the test.

If the check passes, return the machine to service only after completing the entire required pre-use checkout, then stop.

4. Restore the Machine to the Required Test Configuration

Follow the on-screen test instructions and verify that the machine is configured exactly as requested.

Check that:

Do not assume that the same setup applies to every software version or every WATO EX model.

Expected outcome: The test is performed under the conditions expected by the anesthesia system.

5. Inspect the Patient Breathing Circuit

Examine the complete breathing circuit from the inspiratory and expiratory ports to the Y-piece.

Look for:

Reconnect all fittings firmly. Replace questionable disposable tubing, filters, adapters, or test lungs with verified compatible components.

Expected outcome: The external patient circuit is complete, unobstructed, and capable of holding pressure.

Repeat the test. If it passes, document the defective or loose circuit component and stop.

6. Isolate Optional Accessories

Temporarily remove nonessential external accessories that may introduce leakage, including:

Cap or reconnect each port correctly before repeating the test.

Reinstall accessories individually after a successful test to identify the leaking component.

Expected outcome: An accessory-related leak is either identified or ruled out.

7. Check the Y-Piece Test Seal

When the system instructs the operator to occlude the Y-piece:

Expected outcome: The patient connection is completely sealed during the pressure-holding portion of the test.

Repeat the check. If it passes, the previous failure was caused by inadequate Y-piece occlusion.

8. Inspect the Breathing-System Assembly

Without performing internal disassembly, visually confirm that accessible breathing-system components are properly installed.

Check:

A canister, valve cover, or breathing-system component that appears installed but is not fully seated can create a significant leak.

Expected outcome: All user-accessible breathing-system components are properly assembled and sealed.

9. Check the Carbon-Dioxide Absorbent Canister

Inspect the absorbent canister for:

Remove and reinstall the canister only according to the approved operator procedure.

Expected outcome: The absorbent assembly forms a complete seal with the breathing system.

Repeat the leak test. If it passes, stop and document the canister installation issue.

10. Inspect the Manual Breathing Bag and Bag Arm

Verify that:

Replace the breathing bag with a verified compatible bag when its condition is questionable.

Expected outcome: The manual ventilation pathway can hold pressure without leakage.

11. Inspect the Vaporizer Installation

Confirm that each installed vaporizer:

Do not operate or test a vaporizer that has visible agent leakage. Mindray instructs users to ensure vaporizer connections are leak-tight and not to use a vaporizer when anesthetic-agent leakage is present.

When permitted by facility policy, remove a questionable vaporizer using the approved operator procedure and repeat the check.

Expected outcome: The vaporizer and mounting manifold are ruled in or out as the source of the leak.

12. Check Flow Sensors and Accessible Sensor Connections

Inspect user-removable flow sensors and their connections for:

Allow wet components to dry or replace them according to facility and manufacturer procedures. Do not insert tools or compressed air into sensor elements.

Expected outcome: Flow sensors are installed correctly and are suitable for the system check.

13. Repeat the Complete Pre-Use Check

After correcting all identified external issues:

Mindray provides automatic checking functions intended to verify multiple anesthesia-system components and display corrective guidance when a check fails.

Expected outcome: The WATO EX Series completes its required checkout without unresolved failures.

Do not return the machine to clinical service based solely on an isolated test passing if the complete pre-use check remains unsuccessful.

If the Problem Persists

If the pre-use check or leak test continues to fail after the circuit, accessories, breathing-system assembly, canister, bag, vaporizers, supplies, sensors, and required test configuration have been verified, common external causes have been ruled out.

The remaining problem may involve an internal valve, manifold, pressure-control component, pneumatic pathway, sensor, ventilator assembly, or internal seal.

The device should be:

Do not bypass a failed checkout, suppress the result, or return the machine to service because ventilation appears normal during a brief functional test. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a failed anesthesia-machine leak test on an active patient. Move the patient to another verified machine first and perform testing only under safe, controlled conditions.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Mindray WATO EX Series failed the pre-use breathing-system leak test before the scheduled procedure."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the carbon-dioxide absorbent canister incompletely latched, allowing the breathing system to lose pressure during the automatic leak test."

Resolution

What action was taken.

Example:
"Removed and correctly reinstalled the absorbent canister, inspected the patient circuit and accessories, and completed the full pre-use checkout successfully."

Helpful Details to Include (If Known)

Final Thought

A failed pre-use or leak test must be treated as a patient-safety issue. Check external configuration and sealing points logically, repeat the complete checkout, and escalate when the machine cannot demonstrate reliable operation. Accurate CCR documentation preserves what was reported, what was found, and why the device was returned to service or removed for repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide