Mizuho OSI Trios Surgical Table System

Power or Battery Failure

On this page

Asset Type

Surgical Table

Manufacturer

Mizuho OSI

Model

Trios Surgical Table System

What This Guide Helps With

Identifies causes of power or battery failure, including power source, cord, battery, and connection issues before assuming internal electronics failure.

Step-by-Step Troubleshooting

Ensure Patient Safety

Verify External Power Source

Inspect Battery Status

Swap or Recharge Battery

Check Control Panel/Hand Pendant Response

Inspect Fuses and Circuit Breakers

If the Problem Persists

External power, cord, battery, and fuses have been verified. Device may have internal power supply or battery management failure. Remove the table from service, label as Out of Service, and send for bench evaluation or manufacturer repair.

Clinical Use Tip

Never attempt troubleshooting while a patient is on the table. Move the patient to a fully functional table before proceeding. Only Clinical Engineering personnel should handle battery replacements or internal checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Table will not power on; battery does not charge."

Cause

What was observed during troubleshooting.

Example:
"Battery pack faulty or external power source interrupted."

Resolution

What action was taken.

Example:
"Tested outlet and power cord; swapped battery with known good pack; table powers on."

Helpful Details to Include (If Known)

Final Thought

Proper troubleshooting prioritizes patient safety and external factors first. Systematic checks of power sources, battery, and connections prevent unnecessary internal repairs and maintain clinical workflow. Document each step clearly for accountability and trend analysis.

That is successful troubleshooting.