On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots unresponsive controls, abnormal display operation, or missing alarm audio caused by power, contamination, damage, configuration, or hardware failure.
Step-by-Step Troubleshooting
1. Protect the Patient
Immediately remove the pump from active patient therapy if controls cannot be operated reliably, critical information cannot be read, or alarm audio cannot be heard. Transfer therapy to another verified device.
Expected outcome: No patient remains dependent on a pump with unreliable controls, display, or alarm annunciation.
2. Confirm the Exact Failure
Determine which keys do not respond, whether the display is blank, dim, incomplete, frozen, or intermittent, and whether alarm audio is absent, weak, distorted, or continuous. Note whether the fault occurs at startup or during operation.
Expected outcome: The problem is clearly categorized and reproducible when possible.
3. Verify Stable Power
Install known-good C-cell batteries or use an approved external power source as appropriate. Confirm the pump remains powered without resets or flickering.
Expected outcome: A stable power source is established. If the abnormal behavior disappears after correcting the power source and functional testing passes, troubleshooting can stop.
4. Inspect for External Damage or Contamination
Examine the keypad, display window, housing, speaker openings, and surrounding surfaces for cracks, impact marks, fluid intrusion, residue, adhesive buildup, or other contamination. Do not allow liquids to enter the housing during cleaning.
Expected outcome: External surfaces are clean, dry, and mechanically intact. Evidence of liquid intrusion or significant impact requires removal from service.
5. Check Keypad Response
With the pump in a safe nonpatient state, operate each normally accessible control needed for routine function. Do not repeatedly force a stuck key.
Expected outcome: Controls respond consistently to normal pressure without sticking or repeated activation. Any intermittent or stuck control requires further evaluation.
6. Evaluate Display Stability
Observe the display through startup and normal menu or operating transitions. Look for missing information, intermittent blanking, unreadable segments, flickering, or frozen content.
Expected outcome: Information remains readable and updates appropriately. If the display is unreliable, the pump must remain out of service.
7. Verify Alarm Audio
Using an approved nonpatient test method, generate an appropriate alarm condition and verify that audible annunciation is present and clearly detectable. Do not rely solely on a startup sound or button tone.
Expected outcome: Alarm audio operates consistently during an actual test condition. If alarm audio is missing, weak, or intermittent, remove the pump from service.
8. Check Observable Audio Settings
Confirm any user-accessible audio settings are appropriate for the institutional configuration. Do not bypass minimum safety settings, enter unauthorized service menus, or alter protected configuration solely to silence or amplify a suspected fault.
Expected outcome: The complaint is not caused by an obvious authorized setting. If correcting an approved setting restores expected operation, proceed to final testing.
9. Perform Complete Functional Verification
Verify keypad operation, display readability, alarm indicators, alarm audio, power stability, and normal pumping functions using applicable approved procedures.
Expected outcome: All controls and annunciation functions operate reliably. If every required test passes, troubleshooting can stop.
10. Escalate Any Persistent Control, Display, or Audio Fault
If a key remains unresponsive, the display is unreliable, or alarm audio is absent or intermittent, stop external troubleshooting. Do not open the pump for board-level troubleshooting or speaker repair without appropriate service authorization.
Expected outcome: A pump with an unresolved human-interface or alarm-safety fault is kept out of patient care.
If the Problem Persists
Power-source issues, accessible contamination, external damage, observable settings, and normal user controls have been evaluated. Persistent problems may involve the keypad assembly, display circuitry, alarm audio system, internal connections, electronics, or configuration requiring service-level access.
Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. Use manufacturer documentation and approved test equipment. Repairs and protected configuration changes should be completed only by qualified personnel. Verify all alarms, controls, display functions, pumping performance, and other required return-to-service criteria before clinical use.
Knowing when a control or alarm-annunciation problem makes a pump unsafe is proper troubleshooting.
Clinical Use Tip
A pump with silent or unreliable alarm audio should not be returned to patient use simply because the display still shows alarm information.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Nursing reported the CURLIN 6000 alarmed visually but produced no audible alarm during use."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering reproduced the missing alarm audio with a nonpatient test condition despite stable power and normal external settings."
Resolution
What action was taken.
Example:
"Removed the pump from service, labeled it Out of Service, and sent it for qualified bench repair and complete return-to-service testing."
Helpful Details to Include (If Known)
- Keys affected
- Display symptoms
- Alarm audio symptoms
- Alarm indicator behavior
- Battery and external-power results
- Evidence of impact or fluid exposure
- Contamination observed
- Whether the fault was intermittent
- Observable audio settings
- Alarm test result
- Final disposition
- Final device status
Final Thought
Controls, displays, and alarm audio are safety-critical functions; verify power and external causes first, but remove the pump from service whenever reliable operation cannot be demonstrated.
That is successful troubleshooting.