Newport HT70 Plus

Flow Sensor Calibration Errors

On this page

Asset Type

Ventilator

Manufacturer

Newport

Model

HT70 Plus

What This Guide Helps With

Flow sensor calibration failures, inaccurate readings, or alarms caused by external factors such as contamination, setup issues, or environmental conditions.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient ventilation continues safely
Why it matters: Calibration errors can lead to inaccurate tidal volume delivery

Verify the Reported Issue

Expected: Error is reproducible
Why it matters: Confirms a true calibration issue vs user workflow error

Check Circuit and Patient Setup (Most Common Cause)

Expected: Open, unobstructed circuit
Why it matters: Calibration requires an open system with no resistance

Inspect Flow Sensor and Connections

Expected: Sensor is clean, dry, and properly connected
Why it matters: Poor connection or contamination directly affects calibration

Check for Moisture or Contamination (Very Common)

Expected: Dry, clean airflow path
Why it matters: Moisture alters airflow measurement and causes calibration failure

Ensure Proper Calibration Conditions

Expected: Calibration runs without interference
Why it matters: External airflow or resistance will cause failure

Power Cycle the Ventilator

Expected: Calibration completes successfully
Why it matters: Clears temporary software or sensor communication issues

Verify Use of Correct Accessories

Expected: Compatible components installed
Why it matters: Non-approved accessories can cause inaccurate flow readings

Attempt Calibration with Known Good Setup

Expected: Calibration passes with known good components
Why it matters: Helps isolate whether issue is accessory vs device

If the Problem Persists

If calibration continues to fail after verifying circuit, sensor condition, and setup, the issue is likely internal (flow sensor module, internal pneumatics, or control board).

Remove the ventilator from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never attempt flow calibration while connected to a patient. Always transition the patient to another ventilator before troubleshooting to ensure therapy continuity.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator displaying flow sensor calibration error and failing calibration."

Cause

What was observed during troubleshooting.

Example:
"Moisture found in flow sensor and circuit causing inaccurate readings."

Resolution

What action was taken.

Example:
"Replaced circuit and dried sensor; calibration completed successfully."

Helpful Details to Include (If Known)

Final Thought

Flow calibration issues are most often caused by external factors like moisture, circuit resistance, or setup errors. Always rule these out before assuming internal failure. Patient safety and proper escalation are critical, and clear documentation ensures accurate follow-up and trend tracking.

That is successful troubleshooting.