Newport HT70 Plus

Pre-Use Check Failures

On this page

Asset Type

Ventilator

Manufacturer

Newport

Model

HT70 Plus

What This Guide Helps With

Pre-use check failures due to circuit setup, leaks, sensors, or configuration issues preventing the ventilator from passing readiness testing.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Expected: Patient is safely supported on another device

Why it matters: Pre-use check failures indicate the ventilator may not deliver safe or accurate ventilation

2. Verify the Reported Failure

Expected: Failure is repeatable and identifiable

Why it matters: Identifies whether the issue is circuit-related, sensor-related, or system-related

3. Inspect Patient Circuit Setup (Most Common Cause)

Expected: Properly assembled and secure circuit

Why it matters: Incorrect or loose setup frequently causes pre-use check failures

4. Check for Circuit Leaks

Expected: Leak-free system

Why it matters: Even small leaks will cause failure during leak and pressure tests

5. Verify Flow Sensor Installation and Condition

Expected: Proper airflow measurement

Why it matters: Flow sensor issues commonly trigger calibration or pre-check failures

6. Check Exhalation Valve Function (if applicable)

Expected: Valve operates freely

Why it matters: Valve malfunction affects pressure and leak testing

7. Confirm Proper Test Setup (No Patient Connected)

Expected: Proper test environment

Why it matters: Incorrect test setup will cause false failures

8. Check Internal Settings and Mode Configuration

Expected: Settings aligned with test requirements

Why it matters: Incorrect configuration can prevent successful test completion

9. Power Cycle the Ventilator

Expected: System resets and test may pass

Why it matters: Clears temporary software or initialization issues

If the Problem Persists

If all external causes (circuit, sensor, setup, and configuration) have been ruled out, the issue is likely internal (pressure sensors, flow measurement system, or control board). Remove the ventilator from service, label it Out of Service, and send for repair or bench evaluation. Knowing when to stop prevents unnecessary risk and ensures proper escalation.

Clinical Use Tip

Never attempt troubleshooting or pre-use checks while a patient is connected. Always verify ventilator readiness before clinical use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator failed pre-use check and would not pass system test."

Cause

What was observed during troubleshooting.

Example:
"Leak identified in patient circuit due to loose connection at flow sensor."

Resolution

What action was taken.

Example:
"Reseated connections, replaced circuit, and confirmed unit passed pre-use check successfully."

Helpful Details to Include (If Known)

Final Thought

Pre-use check failures are often caused by simple external setup issues rather than internal faults. A structured approach—starting with the circuit and accessories—ensures safe, efficient troubleshooting. Escalate appropriately when needed, and always prioritize patient safety and clear documentation.

That is successful troubleshooting.