On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide assists Clinical Engineering in troubleshooting alarm issues on Nihon Kohden BSM-6000 series patient monitors. Common problems include alarms not sounding, sounding intermittently, sounding at incorrect volumes, or appearing inconsistent with configured limits. Most issues are external, configuration-based, or related to peripheral connections rather than internal electronics.
Step-by-Step Troubleshooting
Verify Power and Battery
- Confirm the monitor is plugged in and turned on.
- Check battery level if running on battery; low battery can disable alarms.
- Why: Alarm circuits may be disabled in low-power conditions.
Check Volume Settings
- Access the monitor’s alarm settings menu.
- Ensure master volume and individual alarm volumes are not muted or set too low.
- Why: Simple misconfiguration is a frequent cause of silent alarms.
Verify Alarm Limits
- Review configured alarm thresholds for each vital sign (HR, SpO₂, NIBP, etc.).
- Ensure values are within realistic ranges to trigger an alarm.
- Why: If limits are set outside expected patient values, alarms will not activate.
Inspect External Connections
- Confirm ECG leads, SpO₂ sensors, NIBP cuffs, and other probes are connected properly.
- Look for disconnected or damaged cables.
- Why: Alarms often rely on valid input signals; disconnected sensors can suppress alarms.
Check Audio Output
- Ensure the internal speaker is unobstructed and functional.
- If an external speaker is connected, verify connections and settings.
- Why: Sound may be routed incorrectly or blocked by physical obstructions.
Perform a System Reset
- Power cycle the monitor following manufacturer instructions.
- Reset user-configurable settings to default if necessary.
- Why: Software glitches or temporary misconfigurations can prevent alarms from sounding.
If the Problem Persists
All external connections, power, and configuration checks have been completed.
The issue is likely internal (speaker module, alarm circuitry, or firmware).
- Remove the device from service.
- Label it Out of Service.
- Send for repair or bench evaluation.
- Avoid further troubleshooting on active patients. Knowing when to escalate is proper troubleshooting.
Clinical Use Tip
- Never rely on a monitor with a non-functioning alarm for patient care.
- Move the patient to another monitored device before performing troubleshooting.
- Always ensure alarms are audible and configured properly prior to patient use.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
“Monitor alarm not sounding during abnormal vitals.”
Cause
What was observed during troubleshooting.
Example:
“Volume set to minimum and NIBP cuff disconnected; master alarm muted.”
Resolution
What action was taken.
Example:
“Adjusted volume settings, reconnected NIBP cuff, verified alarm function with test signal.”
Helpful Details to Include
- Power source tested
- Battery level
- Alarm volume settings reviewed
- Sensor/probe connections verified
- Alarm limits checked
- Audible speaker checked
- Software reset performed
- Final device status
Final Thought
Patient safety relies on functioning alarms. A logical, stepwise approach prevents unnecessary disassembly and ensures proper escalation when issues are internal. Documenting actions clearly supports accountability and continuous quality improvement.
That is successful troubleshooting.