On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide assists Clinical Engineering in troubleshooting NIBP (non-invasive blood pressure) measurement errors or failures on Nihon Kohden BSM-6000 series monitors. Common issues include error messages during measurement, unusually high or low readings, or the inability to initiate NIBP measurement. Most issues are external—related to cuffs, tubing, or connections—before considering internal device faults.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Confirm the patient is not dependent on the monitor for real-time monitoring during troubleshooting.
- Move the patient to an alternative monitor if necessary.
Check NIBP Cuff Application
- Verify the cuff is the correct size for the patient. An undersized or oversized cuff can cause errors.
- Ensure the cuff is positioned correctly on the limb (usually upper arm, at heart level) and wrapped snugly without excessive tightness.
- Inspect for kinks, folds, or damage in the cuff material.
Inspect Tubing and Connections
- Examine the NIBP tubing for cracks, leaks, or disconnections.
- Confirm the tubing is securely connected to both the cuff and the monitor’s NIBP port.
- If possible, swap tubing with a known working set to rule out tubing failure.
Check for Patient Movement or Interference
- Ensure the patient is relaxed and not moving the limb during measurement. Movement can cause failed readings or errors.
- Remove tight clothing or jewelry that may compress the cuff during measurement.
Verify Monitor Settings and Measurement Mode
- Confirm the NIBP measurement mode (automatic, manual, continuous) is correctly configured.
- Ensure alarm thresholds and intervals are set appropriately.
- Restart the monitor if software glitches are suspected.
Observe Monitor Indicators
- Watch for error codes displayed on the monitor. Reference the user guide for error code definitions (e.g., cuff off, excessive movement, low battery).
- Check for unusual sounds, flashing indicators, or error messages that may indicate hardware issues.
If the Problem Persists
After completing all external checks, persistent NIBP failures may indicate internal issues such as pump failure, sensor malfunction, or electronic board fault.
- Remove the monitor from service.
- Label it Out of Service.
- Send for repair or bench evaluation.
- Clinical Engineering should document findings and stop further troubleshooting on active patients.
Clinical Use Tip
- Never attempt NIBP troubleshooting on a patient who is critically dependent on real-time BP readings.
- Always have an alternative monitoring method available while addressing errors.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
“User reports NIBP measurement errors or inability to initiate measurement.”
Cause
What was observed during troubleshooting.
Example:
“Cuff was found loosely wrapped and tubing slightly kinked, causing failed readings.”
Resolution
What action was taken.
Example:
“Repositioned cuff, ensured proper tubing connection, and performed test measurement. Device function restored.”
Helpful Details to Include
- Outlet tested: N/A for NIBP, but verify monitor powers on.
- Cables swapped: Tubing swapped with a known good set.
- Alarm behavior: Monitor displayed error code “Cuff Off” during initial test.
- Indicator lights: NIBP indicator flashed red during error.
- Unusual sounds: Pump attempted inflation but released quickly.
- Final device status: Working after cuff/tubing adjustment; patient monitored safely.
Final Thought
Proper NIBP troubleshooting emphasizes patient safety, external factor verification, and logical escalation. Documenting CCRs ensures accountability and aids future maintenance, while preventing unnecessary vendor costs. Only escalate to repair after ruling out all external and easily fixable causes.
That is successful troubleshooting.