Nihon Kohden NKV-330

Alarm: AC Power Loss/Backup Battery In Use

On this page

Asset Type

Ventilator

Manufacturer

Nihon Kohden

Model

NKV-330

What This Guide Helps With

This guide helps troubleshoot the Nihon Kohden NKV-330 alarm AC Power Loss/Backup Battery In Use. This alarm indicates that AC power and main battery power are lost, and the ventilator is operating on backup battery power.

This condition should be treated as urgent because normal power redundancy is compromised and remaining backup battery runtime may be limited.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Connect the Ventilator to AC Power Immediately

Verify the Outlet Has Power

Inspect the Power Cord and AC Connection

Check the Main Battery Status

Check for Recent Transport or Power Change

Confirm Alarm Clears and Power Status Is Stable

Stop When Resolved

If the Problem Persists

If the alarm continues after confirming a known good AC outlet, secure power cord connection, and known good main battery, external and common causes have been ruled out.

Persistent AC Power Loss/Backup Battery In Use conditions may indicate an internal power input, battery charging, battery detection, or power management failure.

Clinical Use Tip

Never troubleshoot this alarm casually while the ventilator is supporting an active patient. Backup battery operation means power redundancy is compromised. Move the patient to a backup ventilator or restore stable power immediately if the alarm cannot be cleared quickly.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nihon Kohden NKV-330 ventilator displayed Alarm: AC Power Loss/Backup Battery In Use. Staff reported the ventilator was operating on backup battery power after loss of AC and main battery power."

Cause

What was observed during troubleshooting.

Example:
"Ventilator was connected to a non-powered outlet after transport, and the main battery was depleted. Backup battery alarm activated as expected."

Resolution

What action was taken.

Example:
"Connected ventilator to known working AC emergency-power outlet. Installed fully charged spare main battery. Confirmed AC power restored, main battery recognized, charging status present, and alarm cleared. Monitored operation with no recurrence. Returned device to service."

Helpful Details to Include (If Known)

Final Thought

The AC Power Loss/Backup Battery In Use alarm should be treated as urgent because the ventilator has lost normal AC power and main battery support. Start with patient safety, restore AC power, verify the main battery, and escalate if stable power cannot be confirmed. Clear documentation protects the patient, the clinical team, and the repair process.

That is successful troubleshooting.