Nihon Kohden NKV-330

Alarm: AC Power Loss/Main Battery In Use

On this page

Asset Type

Ventilator

Manufacturer

Nihon Kohden

Model

NKV-330

What This Guide Helps With

This guide helps troubleshoot the AC Power Loss/Main Battery In Use alarm on the Nihon Kohden NKV-330 ventilator. This alarm means AC power is lost and the ventilator is operating on main battery power.

The ventilator may still be supporting ventilation, but battery runtime is limited. The priority is to confirm patient safety, restore AC power, check external power sources, and remove the ventilator from service if it cannot recognize verified AC power.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check the Wall Outlet

4. Check the Power Cord Connection

5. Restore AC Power and Confirm the Alarm Clears

6. Check Remaining Main Battery Power

7. Check for a Facility Power Problem

8. Isolate Device Versus External Power Issue

If the Problem Persists

If the AC Power Loss/Main Battery In Use alarm remains active after confirming the wall outlet, power cord, and external AC source are good, external and common causes have been ruled out.

A persistent alarm under verified AC power conditions may indicate an internal power input, AC detection, charging, or power supply problem.

Clinical Use Tip

Never troubleshoot power-source alarms on an active patient without clinical support. Move the patient to a backup ventilator first if AC power cannot be restored quickly or if the ventilator power condition is unreliable.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nihon Kohden NKV-330 ventilator displayed AC Power Loss/Main Battery In Use alarm and staff reported the ventilator was operating on main battery power."

Cause

What was observed during troubleshooting.

Example:
"AC power was not being detected by the ventilator. Outlet and power cord were checked, and the issue was related to the external power source or cord connection."

Resolution

What action was taken.

Example:
"Verified patient safety, checked remaining battery level, inspected power cord and AC connections, tested outlet, restored AC power, and confirmed the alarm cleared. Ventilator was returned to service after normal operation was verified."

Helpful Details to Include (If Known)

Final Thought

The AC Power Loss/Main Battery In Use alarm should be treated as a patient safety and power reliability issue first. Confirm ventilation, preserve battery runtime, check the outlet and power cord, and restore AC power as soon as possible. If the ventilator does not recognize verified AC power, remove it from service and escalate for repair. Clear documentation helps separate facility power issues from true ventilator power-system failures.

That is successful troubleshooting.