Nihon Kohden NKV-330

Alarm: High PEEP

On this page

Asset Type

Ventilator

Manufacturer

Nihon Kohden

Model

NKV-330

What This Guide Helps With

This guide helps troubleshoot the High PEEP alarm on a Nihon Kohden NKV-330 ventilator. This alarm indicates measured PEEP is more than 5 hPa above the EPAP/CPAP setting. Possible causes include patient airway resistance, circuit restriction, water accumulation, blocked expiratory flow, incorrect settings, pressure or flow sensor line problems, calibration issues, or a ventilator delivery fault.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check the Patient and Airway

Check Control and Alarm Settings

Inspect the Patient Interface

Inspect the Breathing Circuit

Check the Exhalation Port

Check the Pressure Sensor Line

Check the Flow Sensor Line

Calibrate the Pressure Sensor and Breathing Circuit

Recheck Operation After Corrective Action

If the Problem Persists

If the High PEEP alarm continues after checking the patient, settings, patient interface, breathing circuit, exhalation port, pressure sensor line, flow sensor line, and calibration, external and common causes have been ruled out.

A persistent High PEEP alarm may indicate an internal breath delivery or pressure-control problem. The CD-358P breath delivery unit may be faulty.

Clinical Use Tip

Never troubleshoot a High PEEP alarm on an active patient if ventilation safety is uncertain. Move the patient to a backup ventilator or approved alternate ventilation method first, then continue equipment troubleshooting on the removed device.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"NKV-330 ventilator reported High PEEP alarm. Measured PEEP was more than 5 hPa above the EPAP/CPAP setting during ventilation."

Cause

What was observed during troubleshooting.

Example:
"Water droplets and partial blockage were found in the expiratory circuit and pressure sensor line, causing elevated measured PEEP."

Resolution

What action was taken.

Example:
"Removed affected circuit components, replaced contaminated tubing, verified exhalation port was clear, completed pressure sensor and breathing circuit calibration, and confirmed the High PEEP alarm did not return during testing."

Helpful Details to Include (If Known)

Final Thought

A High PEEP alarm should be treated as a real pressure-retention concern until proven otherwise. Start with patient safety, confirm the measured pressure condition, check the circuit and sensing lines carefully, then calibrate and test. If the alarm persists after external causes are ruled out, remove the ventilator from service and escalate appropriately. Clear CCR documentation protects the patient, the technician, and the repair process.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide