NxStage System One

Air Detector or Venous Pressure Sensor Alarms

On this page

Asset Type

Hemodialysis (HD) Machine

Manufacturer

NxStage

Model

System One

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where an NxStage System One HD machine triggers air detector alarms or abnormal venous pressure readings. These alarms typically indicate external causes such as tubing kinks, disconnections, or sensor blockages. Internal sensor failure is less common but possible. The focus is on simple, safe, external checks before escalating for repair.

Step-by-Step Troubleshooting

Ensure Patient Safety

Inspect Tubing and Connections

Check for Air in the Circuit

Inspect the Pressure Transducers

Examine the Air Detector Chamber

Swap Components if Needed

Check Machine Settings

If the Problem Persists

External and easily verifiable issues have been ruled out. The issue may be due to internal sensor failure or a board fault.

Next Steps:

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Machine triggered air detector and venous pressure alarms during treatment.”

Cause

What was observed during troubleshooting.

Example:
“Arterial line was kinked; air detector chamber properly seated but clogged with residual fluid.”

Resolution

What action was taken.

Example:
“Removed kink in tubing; replaced air detector chamber; tested machine with a dummy circuit. Alarms cleared and machine returned to service.”

Helpful Details to Include

Final Thought

Patient safety is paramount. Following a logical, stepwise troubleshooting approach ensures external causes are addressed first, minimizing unnecessary downtime or risk. Proper documentation using the CCR method provides accountability and clarity for any future service needs.

That is successful troubleshooting.