NxStage System One

Flow Imbalance or Dialysate Preparation Faults

On this page

Asset Type

Hemodialysis (HD) Machine

Manufacturer

NxStage

Model

System One

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting flow imbalance or dialysate preparation faults on an NxStage System One HD machine. These alarms indicate that the machine has detected irregular dialysate flow, improper mixing, or other preparation issues. Most commonly, these are caused by external factors such as kinks in tubing, improper cartridge installation, or blocked filters rather than internal electronic failures.

Step-by-Step Troubleshooting

Verify Machine Power and Startup

Why: Unstable power or incomplete initialization can trigger false flow or dialysate alarms.

Inspect Dialysate Cartridge

Why: Improper cartridge installation can impede fluid flow and mixing.

Examine Tubing and Connections

Why: Flow imbalances often result from restricted or misconnected tubing.

Check Dialysate Concentrate and Filters

Why: Contaminated or misconnected filters can reduce flow or cause preparation faults.

Prime the System and Observe Flow

Why: Priming helps confirm whether flow issues persist after initial setup.

Review Alarm History and Machine Logs

If the Problem Persists

External causes have been largely ruled out.

Device should be:

Reminder: Escalating before attempting internal repair ensures patient safety and preserves warranty.

Clinical Use Tip

Never troubleshoot flow or dialysate faults while the machine is connected to a patient.

Move the patient to a backup dialysis machine before performing troubleshooting steps.

Ensure all safety alarms are acknowledged and that the treatment environment remains safe.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Flow imbalance” or “Dialysate preparation fault” during patient treatment.

Cause

What was observed during troubleshooting.

Example:
Observed kinked dialysate tubing and loosely seated cartridge; dialysate filters inspected and clear.

Resolution

What action was taken.

Example:
Corrected cartridge seating, straightened tubing, and verified proper priming. Flow readings normalized. Machine returned to service.

Helpful Details to Include

Final Thought

Patient safety is paramount. Start troubleshooting with external, easily verifiable issues and avoid internal disassembly unless directed by the manufacturer. Proper documentation and timely escalation preserve both safety and machine integrity.

That is successful troubleshooting.