NxStage System One

Pump Module or Flow Controller Failure

On this page

Asset Type

Hemodialysis (HD) Machine

Manufacturer

NxStage

Model

System One

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where an NxStage System One HD machine reports pump module or flow controller failures. These alarms typically indicate that the machine cannot maintain proper blood or dialysate flow due to sensor issues, pump malfunctions, or connection faults. External causes should be verified first before assuming internal failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check Power and Connections

Inspect Pump Tubing and Lines

Verify Pump Module Installation

Check Flow Sensors

Observe Alarm Patterns and Logs

If the Problem Persists

External factors have been ruled out. The pump module or flow controller is likely internally faulty.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump module failure alarm displayed during startup.”

Cause

What was observed during troubleshooting.

Example:
“Observed that the pump module was properly seated but the controller repeatedly failed self-test. No external tubing or sensor issues were found.”

Resolution

What action was taken.

Example:
“Machine removed from service, labeled Out of Service, and sent for vendor repair.”

Helpful Details to Include

Final Thought

Patient safety is the highest priority. Clinical Engineering troubleshooting should follow a logical, stepwise process starting with external checks. Recognizing when to escalate to repair prevents further risk and ensures reliable dialysis treatment. Proper documentation supports regulatory compliance and future reference.

That is successful troubleshooting.