Olympus EVIS EXERA III

System Not Powering On or Random Shutdowns

On this page

Asset Type

Endoscopic / OR Camera System

Manufacturer

Olympus

Model

EVIS EXERA III

What This Guide Helps With

Addresses no power or unexpected shutdowns caused by power supply issues, loose connections, overheating, or external accessory faults.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Check AC Power Source (Most Common Cause)

Inspect Power Cord and Connections

Check System Power Switch and Indicators

Inspect for Overheating Conditions

Disconnect External Accessories

Check Power Distribution (If in Equipment Tower)

Attempt Power Cycle Reset

If the Problem Persists

If all external power sources, cables, and environmental conditions have been verified, the issue is likely internal (power supply, internal board, or thermal sensor failure).

Remove the device from service

Label as Out of Service

Send for repair or further bench evaluation

Knowing when to stop prevents unnecessary risk and is part of proper troubleshooting.

Clinical Use Tip

Do not troubleshoot power issues while the system is in active clinical use. Always move the procedure to a functioning backup system before beginning evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"EVIS EXERA III system would not power on and intermittently shut down during use."

Cause

What was observed during troubleshooting.

Example:
"Loose AC power cord connection at system inlet causing intermittent power loss."

Resolution

What action was taken.

Example:
"Secured power connection and verified stable operation through power cycle testing."

Helpful Details to Include (If Known)

Final Thought

Power issues should always be approached from the outside in—starting with simple, verifiable causes before assuming internal failure. Patient safety comes first, especially with visualization-dependent procedures. Proper escalation and clear documentation ensure reliability and accountability in Clinical Engineering workflows.

That is successful troubleshooting.