Air/Water Pump or CO2 Insufflation Is Unavailable

On this page

Asset Type

Endoscopic / OR Camera System

Manufacturer

Olympus

Model

EVIS EXERA III

What This Guide Helps With

Troubleshoots unavailable air, water, or CO2 delivery caused by connections, tubing, bottles, accessories, supply sources, controls, or configuration.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain the Procedure Safely

If air, water, or CO2 delivery is required during an active procedure and the function is unavailable or unreliable, clinical staff must move to an approved alternate method or verified system before troubleshooting.

Do not disconnect insufflation or fluid-delivery components while the patient depends on them.

Expected outcome: Patient care continues using a dependable alternative while the affected system is evaluated.

2. Confirm Which Function Is Unavailable

Determine whether the complaint involves:

Confirm whether the problem began after tubing, water-bottle, CO2-source, scope, or accessory replacement.

Expected outcome: The affected function and operating conditions are clearly identified.

3. Verify System Power and Component Availability

Confirm the required processor, light source, pump, insufflator, and related equipment are powered and have completed normal startup.

Check power cords, outlet connections, and external power indicators.

Expected outcome: All required components are powered and available. If the problem was caused by a disconnected or unpowered external component, correct it, verify delivery, and stop troubleshooting.

4. Inspect Air/Water and Insufflation Connections

Inspect all externally accessible tubing and connections associated with the affected function.

Look for:

Do not force incompatible fittings.

Expected outcome: Tubing is properly connected, unobstructed, and undamaged. If correcting the connection restores normal delivery, perform final verification and stop troubleshooting.

5. Check the Water Bottle and Related Accessories

If water delivery is affected, inspect the installed water bottle and connecting components.

Verify:

Expected outcome: The water supply assembly is properly installed and free of obvious external defects.

6. Verify the CO2 Supply Path

If CO2 insufflation is unavailable, verify the approved external CO2 supply and associated insufflation equipment are available and properly connected.

Check for:

Do not alter regulated gas systems or pressure settings outside authorized procedures.

Expected outcome: The external CO2 supply path is available and correctly connected.

7. Inspect the Scope Connection and Applicable Valves

Verify the scope is correctly connected and that removable external valves or accessories associated with air/water delivery are installed properly.

Inspect for visible contamination, damage, incorrect assembly, or blockage.

Only perform cleaning or replacement permitted by approved scope-handling procedures.

Expected outcome: The scope-side external delivery components are correctly installed and unobstructed.

8. Test With Known-Good Compatible Tubing or Accessories

When available, substitute known-good compatible tubing, water-bottle components, valves, scope accessories, or another compatible scope one item at a time.

Do not substitute components that are not approved for the system.

Expected outcome: If function returns with a known-good external component, remove the defective accessory from service. If the failure remains, continue troubleshooting.

9. Verify User-Accessible Controls and Configuration

Check whether the affected air/water or CO2 function has been disabled through an accessible control, procedure profile, or connected accessory selection.

Compare settings with a known-good EVIS EXERA III configuration when appropriate.

Do not enter unauthorized service menus.

Expected outcome: Accessible controls and configuration are appropriate for the intended function. If correcting an inadvertent selection restores operation, document it and stop troubleshooting.

10. Perform Controlled Functional Verification

Using an approved bench setup, verify the affected function without a patient connected.

Confirm:

Expected outcome: The required delivery function operates consistently. Troubleshooting can stop when final functional verification is successful.

If the Problem Persists

If air, water, or CO2 remains unavailable after checking power, supplies, tubing, bottles, valves, scope connections, compatible accessories, and accessible controls, common external causes have been ruled out.

The remaining cause may involve an internal pump, control interface, gas-management subsystem, scope communication condition, configuration issue, or another service-level fault.

The affected equipment should be:

Do not bypass delivery controls or perform unauthorized pneumatic or internal repairs. Complete appropriate functional testing before returning the system to service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Maintain the required insufflation or irrigation function with an approved alternate method before removing any affected equipment from an active procedure.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that water delivery was unavailable from the EVIS EXERA III during setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the water-bottle tubing was kinked behind the cart and restricting flow."

Resolution

What action was taken.

Example:
"Corrected the tubing route, verified normal water delivery and secure connections during bench testing, and returned the system to service."

Helpful Details to Include (If Known)

Final Thought

Start with patient continuity, then verify the external supply path, tubing, bottles, scope accessories, controls, and configuration before assuming an internal pump or gas-management failure. Confirm normal operation before return to service and clearly document the final finding.

That is successful troubleshooting.