Olympus OER-Pro

Chemical Level or Dosing Errors

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Olympus

Model

OER-Pro

What This Guide Helps With

Troubleshooting chemical level alarms, dosing failures, or disinfectant errors caused by empty containers, sensor issues, or improper connections.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: No risk to patient from improperly disinfected scope
Why it matters: Chemical dosing directly impacts high-level disinfection effectiveness

Verify the Reported Error and Behavior

Expected: Clear identification of which chemical system is affected
Why it matters: Narrows troubleshooting to the correct supply system

Check Chemical Container Levels (Most Common Cause)

Expected: Containers have adequate fluid levels
Why it matters: Low or empty containers will trigger dosing failures

Inspect Chemical Pickup Tubing and Connections

Expected: Continuous, airtight fluid path
Why it matters: Air in the line or loose tubing prevents proper dosing

Check Container Caps and Float Sensors

Expected: Sensors move freely and detect fluid levels correctly
Why it matters: False low-level readings can stop dosing even when fluid is present

Verify Correct Chemical Installation and Recognition

Expected: No mismatch or recognition errors
Why it matters: Incorrect or unrecognized chemicals can trigger dosing faults

Check for Air in Lines / Prime System (If Applicable)

Expected: Smooth fluid flow with no air pockets
Why it matters: Air prevents accurate chemical delivery

Restart the System and Reattempt Cycle

Expected: No chemical alarms during operation
Why it matters: Confirms issue was external and resolved

If the Problem Persists

If chemical levels, tubing, and sensors have been verified and errors continue, the issue is likely internal (pump failure, sensor malfunction, or control board issue).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unsafe disinfection conditions and protects patient safety.

Clinical Use Tip

Never bypass or ignore chemical alarms. If proper dosing cannot be confirmed, the scope must not be used. Always move to a functioning AER before continuing reprocessing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"AER reporting chemical level error and unable to start disinfection cycle."

Cause

What was observed during troubleshooting.

Example:
"Disinfectant container empty and pickup tube drawing air."

Resolution

What action was taken.

Example:
"Replaced disinfectant container, primed system, and verified successful cycle completion."

Helpful Details to Include (If Known)

Final Thought

Chemical dosing errors should always be approached from the outside in. Most issues are simple—empty containers or loose tubing—but the impact is critical. Proper escalation protects patients and ensures disinfection integrity.

That is successful troubleshooting.