Outset Medical Tablo

Arterial or Venous Pressure Reading Incorrect or Unstable

On this page

Asset Type

Hemodialysis (HD) Machine

Manufacturer

Outset Medical

Model

Tablo

What This Guide Helps With

Troubleshoot inaccurate or unstable arterial or venous pressure readings caused by tubing, pressure connections, air, clamping, disposable setup, or patient-circuit conditions.

Step-by-Step Troubleshooting

1. Protect the Patient

Do not rely on unstable or clearly incorrect arterial or venous pressure monitoring during dialysis. If the issue occurs during treatment, clinical staff should assess the patient and extracorporeal circuit and follow facility procedure for safely stopping or transferring treatment when required.

Technical troubleshooting should occur off patient whenever device performance is uncertain.

Expected outcome: The patient is not dependent on unreliable pressure monitoring during troubleshooting.

2. Confirm Which Pressure Channel Is Affected

Determine whether the issue involves:

Record the exact reported behavior.

Expected outcome: The affected channel and failure pattern are identified.

3. Inspect the Extracorporeal Tubing Setup

With the machine off patient, inspect the blood tubing and pressure-monitoring portions of the disposable setup for:

Expected outcome: The tubing path is installed correctly and free of obvious restrictions. If correcting an external setup issue stabilizes pressure readings, troubleshooting can stop after verification.

4. Inspect Pressure Connections and Interfaces

Check accessible pressure interfaces and connections for secure seating, contamination, moisture, loose fittings, or incorrectly installed disposable components.

Do not open internal pressure-sensing assemblies.

Expected outcome: Pressure interfaces are clean, dry, properly connected, and mechanically secure.

5. Check for Air or Fluid Where It Should Not Be

Inspect accessible pressure-monitoring portions of the circuit for abnormal air or fluid conditions that may interfere with pressure transmission.

Use normal setup procedures only; do not improvise purging or pressure-source methods.

Expected outcome: The disposable pressure-monitoring path is in the expected setup condition.

6. Replace the Suspect Disposable

If the tubing or pressure interface remains questionable, replace the disposable set with a new approved set and repeat setup.

This is particularly useful when a problem follows one specific tubing set.

Expected outcome: Stable pressure behavior with the replacement disposable indicates the original disposable or setup caused the condition.

7. Observe Pressure Behavior During Normal Off-Patient Testing

Use the normal machine preparation or approved functional test process to observe whether the affected reading responds consistently.

Do not apply unapproved pressure sources or enter unauthorized calibration functions.

Expected outcome: Pressure readings behave consistently and no longer fluctuate abnormally.

8. Check Related Pump and Clamp Conditions

Confirm there is no simultaneous blood-pump, tubing, or clamp problem that could account for abnormal pressure behavior.

A pressure alarm may be the result of an actual flow restriction rather than a failed pressure sensor.

Expected outcome: Other external components that influence circuit pressure are operating normally.

9. Perform Final Functional Verification

Complete the normal pre-treatment verification sequence and confirm arterial and venous pressure monitoring is stable with the approved setup.

Expected outcome: Both pressure channels operate consistently with no unexplained alarm or instability. Troubleshooting can stop.

10. Escalate Persistent Pressure-Monitoring Problems

If the disposable, pressure connections, tubing path, and related external conditions are satisfactory but readings remain unreliable, remove the machine from service.

Expected outcome: Internal sensing, pneumatic interface, signal processing, calibration, or other service-level problems are referred for qualified evaluation.

If the Problem Persists

Common external tubing, disposable, connection, air/fluid, clamping, and flow-related causes have been ruled out. A persistent problem may involve internal pressure sensing, signal processing, pneumatic control, calibration, or associated electronics.

The device should be:

Pressure-monitoring accuracy and applicable dialysis safety functions must be verified before return to service. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Pressure alarms may reflect a real extracorporeal-circuit restriction; confirm the circuit condition before assuming the monitor is inaccurate.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported the Tablo displayed an unstable venous pressure value that changed significantly with minor tubing movement."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the venous pressure connection on the disposable was incompletely seated."

Resolution

What action was taken.

Example:
"Installed the pressure connection correctly, repeated setup, verified stable arterial and venous pressure monitoring, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Unstable dialysis pressure readings require verification of the entire external blood-circuit setup before an internal sensor is blamed. Protect the patient, rule out real flow restrictions and disposable problems, verify stable monitoring after correction, and escalate unresolved sensing issues appropriately.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide