Outset Medical Tablo

Conductivity Out of Range or Dialysate Will Not Stabilize

On this page

Asset Type

Hemodialysis (HD) Machine

Manufacturer

Outset Medical

Model

Tablo

What This Guide Helps With

Troubleshoot unstable dialysate conductivity by checking concentrates, water preparation, consumables, connections, setup, and external conditions before service escalation.

Step-by-Step Troubleshooting

1. Protect the Patient

Do not continue dialysis when dialysate conductivity is unreliable or outside the machine's accepted operating condition. If the issue occurs during treatment, follow the facility's clinical procedure for safely stopping, maintaining, or transferring therapy.

Never bypass a conductivity alarm.

Expected outcome: The patient is not exposed to dialysate of uncertain composition, and the affected machine is available for off-patient troubleshooting.

2. Confirm the Conductivity Complaint

Determine whether conductivity:

Record exact messages and the point in the sequence where the condition appears.

Expected outcome: The failure pattern is documented and reproducible.

3. Verify the Correct Concentrates and Consumables

Confirm that all required concentrates and disposable components are the correct type for the intended treatment and are installed according to normal clinical setup.

Inspect labels, connections, expiration status when applicable, and obvious container damage or contamination.

Do not substitute products that are not approved for the system.

Expected outcome: Correct supplies are present and properly connected. If an incorrect or improperly connected supply is corrected and conductivity stabilizes, troubleshooting can stop after verification.

4. Inspect Concentrate Connections and Fluid Paths

Check accessible concentrate connections for:

Avoid opening internal fluid-path components.

Expected outcome: External concentrate pathways are secure, unobstructed, and properly supplied.

5. Verify Water Preparation Is Normal

Confirm the machine has completed its normal water-preparation process without unresolved alarms. Review whether any recent water inlet or integrated purification problems occurred.

Conductivity instability associated with a water-preparation problem should be addressed as a water-system issue before further dialysis troubleshooting.

Expected outcome: Water preparation is stable and no unresolved water-related condition is present.

6. Check Setup and Configuration

Verify the selected treatment setup and accessible parameters are appropriate for the intended therapy and have not been accidentally changed.

Do not alter restricted calibration or service parameters.

Expected outcome: The machine is configured for the intended treatment without unintended user-level setup changes.

7. Replace Questionable External Supplies With Known-Good Supplies

If a concentrate container, connector, or disposable is suspect, replace it with an approved known-good equivalent when permitted by facility procedure.

Change one variable at a time so the corrective action can be identified.

Expected outcome: Conductivity stabilizes with the known-good supply if the original external component caused the problem. Troubleshooting can stop after successful verification.

8. Repeat the Preparation Sequence

With the machine off patient, repeat the normal preparation process and observe whether conductivity reaches and maintains the system's accepted condition.

Do not bypass or silence recurring safety checks merely to advance the setup.

Expected outcome: Conductivity stabilizes and remains accepted by the system. If so, continue to final verification.

9. Perform Final Functional Verification

Confirm the Tablo completes pre-treatment preparation without conductivity alarms and maintains stable dialysate preparation through the normal verification period.

Expected outcome: No recurrent conductivity alarm occurs and the machine reaches the expected ready state. Troubleshooting can stop.

10. Escalate Persistent Conductivity Instability

If correct concentrates, proper connections, normal water preparation, and appropriate setup are confirmed but conductivity remains unstable, remove the device from service.

Expected outcome: Internal proportioning, sensing, fluid-control, or calibration-related causes are left to qualified service evaluation.

If the Problem Persists

Common external concentrate, connection, consumable, setup, and water-preparation causes have been ruled out. A persistent problem may involve internal sensing, mixing, pumping, fluid-control, calibration, or service-level configuration.

The device should be:

Conductivity and applicable dialysis-fluid performance must be verified before return to service. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Never clear a conductivity alarm and proceed with treatment until the machine has demonstrated stable, acceptable dialysate preparation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Dialysis staff reported the Tablo would not stabilize conductivity during treatment preparation."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the concentrate connection was not completely seated, resulting in inconsistent concentrate delivery."

Resolution

What action was taken.

Example:
"Reseated the approved concentrate connection, repeated preparation, verified stable conductivity with no recurring alarm, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

Conductivity problems must be treated as patient-safety issues. Confirm supplies, connections, water preparation, and setup before assuming an internal fault, and never bypass protective checks. Verify stability after correction, escalate persistent problems, and document the complete event.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide