What This Guide Helps With

Troubleshoots loss of network or IntelliSpace Perinatal communication through external connections, network availability, configuration, ports, infrastructure, and endpoint isolation.

Step-by-Step Troubleshooting

1. Protect the Patient and Establish an Alternate Monitoring Workflow

A local fetal monitor may continue displaying patient data even when network communication is unavailable, but remote surveillance, charting, or data distribution may be affected.

Notify clinical staff of the communication loss and establish the approved alternate monitoring and documentation process before troubleshooting.

Expected outcome: Clinical staff are not relying on an unavailable central or networked monitoring path.

2. Confirm the Scope of the Failure

Determine whether:

Confirm the destination or system that is not receiving data.

Expected outcome: The issue is classified as device-specific, room-specific, or system-wide.

3. Verify Local Monitor Operation

Confirm the FM50 is operating normally at the bedside and displaying expected monitoring information.

If the monitor itself is unstable, restarting, or losing parameters, resolve the local equipment issue separately before treating the problem as network-only.

Expected outcome: The monitor functions normally aside from communication.

4. Inspect External Network Connections

Inspect accessible network and communication cables for:

Reseat connections at the monitor and accessible wall or approved interface points.

Expected outcome: All external communication connections are secure and physically intact.

5. Check Available Link or Communication Indicators

Observe any normal accessible link, communication, or connection indicators on the monitor, network adapter, or approved external interface.

Do not interpret an indicator beyond what is supported by local documentation, but use it to compare the affected setup with a known-working setup.

Expected outcome: Physical network connectivity appears comparable to a functioning station.

6. Substitute a Known-Good External Network Cable

Replace the accessible device-side network cable with a known-good cable.

Do not disturb permanent facility cabling beyond approved Clinical Engineering scope.

Expected outcome: Communication is restored if the original patch cable was defective.

If communication remains stable after cable replacement, complete end-to-end verification and stop troubleshooting.

7. Compare the Network Location

When operationally appropriate, test the affected monitor on a known-working approved network connection, or test a known-good compatible monitor at the affected location.

Coordinate moves and network testing with clinical operations and facility policy.

Expected outcome: The problem follows either the monitor or the room/network location.

8. Verify Authorized Network Configuration

Review the monitor's authorized network configuration and compare it with current site records or a known-good equivalent unit.

Do not change IP addressing, VLAN-related configuration, destination settings, or other network parameters without authorization and validated site information.

Expected outcome: Device configuration matches the intended network environment.

9. Confirm the Complete IntelliSpace Perinatal Path

Verify whether the device appears at the expected receiving system or workstation and whether data from other monitors is arriving normally.

Coordinate with the responsible perinatal application or network support team when the failure extends beyond the bedside connection.

Expected outcome: The fault is isolated to the FM50, network infrastructure, or receiving application.

10. Perform End-to-End Functional Verification

After correcting the identified issue, verify:

Use test data or an approved nonpatient verification method whenever practical.

Expected outcome: Communication from the Avalon FM50 through the intended network path is restored and stable.

If successful, troubleshooting can stop.

11. Escalate Unresolved Communication Failure

If the monitor works on one network location but not another, escalate the affected port, cabling, switch, VLAN, or infrastructure path to the appropriate network team.

If the failure follows the FM50 across known-good connections, remove the monitor from service when network integration is required and route it for service evaluation.

Expected outcome: Responsibility is transferred to the correct technical group with the problem clearly isolated.

If the Problem Persists

External connections, patch cables, location comparison, local monitor operation, and authorized configuration have been checked.

Persistent communication failure may involve the monitor's communication interface, hospital network infrastructure, network configuration, IntelliSpace Perinatal services, or another integration-level condition.

The device should be:

Infrastructure or application failures should be escalated to the appropriate hospital network or IntelliSpace Perinatal support team.

Complete end-to-end communication verification before return to service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Verify the complete monitoring path at the receiving station; a normal bedside display does not confirm that remote fetal monitoring or documentation is functioning.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Labor and delivery staff reported that one Avalon FM50 displayed normally at the bedside but was no longer communicating with IntelliSpace Perinatal."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a defective external network patch cable, while the monitor communicated normally after connection with a known-good cable."

Resolution

What action was taken.

Example:
"Replaced the patch cable and verified stable bedside operation and successful end-to-end communication with the intended IntelliSpace Perinatal destination."

Helpful Details to Include (If Known)

Final Thought

Network troubleshooting should follow the entire communication path from the bedside monitor through external cabling and infrastructure to the receiving application. Preserve clinical monitoring, isolate the affected segment logically, avoid unauthorized network changes, and document the final end-to-end verification.

That is successful troubleshooting.