Philips Brilliance Series

Table Motion or Positioning Errors

On this page

Asset Type

CT Scanner

Manufacturer

Philips

Model

Brilliance Series

What This Guide Helps With

Table movement, indexing, positioning, or height errors caused by obstructions, controls, accessories, power, or positioning system faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not continue scanning if the table is moving unpredictably, not stopping correctly, or positioning inaccurately.

Action:

Expected outcome:
Patient safety is maintained before troubleshooting begins.

If this resolves the immediate safety concern, stop and document the issue.

2. Verify the Reported Table Problem

Confirm the exact behavior before assuming a table failure.

Action:

Expected outcome:
The table motion or positioning problem is clearly identified and reproducible.

3. Confirm the Table Is Clear of the Patient and Staff

Physical obstruction is one of the first things to rule out.

Action:

Expected outcome:
Table travel path is clear.

If the table moves normally after clearing the obstruction, return the unit to service after verifying safe operation and documenting the finding.

4. Check Emergency Stop and Motion Inhibit Conditions

A CT table may be prevented from moving if an emergency stop or safety interlock is active.

Action:

Expected outcome:
No emergency stop or safety inhibit is active.

If resetting a known activated emergency stop restores table motion, verify full table operation before returning to service.

5. Inspect Table Controls and Gantry Control Panels

A stuck or damaged control can cause no movement, intermittent movement, or unexpected behavior.

Action:

Expected outcome:
Controls respond consistently and do not appear stuck, damaged, or contaminated.

If one control location fails but another works normally, document the failed control location and remove the scanner from service if safe operation cannot be guaranteed.

6. Check Table Position Limits

The table may stop moving if it has reached a mechanical or programmed travel limit.

Action:

Expected outcome:
The table either moves away from the limit or confirms a travel-limit condition.

If the table was simply at a travel limit and normal movement resumes, verify full operation and document the finding.

7. Verify Patient Weight and Table Load

Excessive load or uneven loading can affect table lift and positioning accuracy.

Action:

Expected outcome:
The table is not overloaded or unevenly loaded.

If table motion returns after correcting load conditions, document the cause and remind staff to follow table weight and positioning limits.

8. Inspect Tabletop, Cradle, and Accessories

Misaligned or loose accessories can interfere with table motion or indexing.

Action:

Expected outcome:
No accessory or tabletop issue is interfering with table travel.

If an accessory was causing the issue, remove it from use and document the correction.

9. Check for Abnormal Noise, Binding, or Vibration

Mechanical resistance can indicate a table drive, rail, bearing, lift, or alignment issue.

Action:

Expected outcome:
Table motion is smooth and controlled.

If abnormal noise, binding, or vibration is present, remove the scanner from service and escalate for repair.

10. Verify Console Position Display and Table Position Accuracy

A positioning mismatch can affect scan planning and image location accuracy.

Action:

Expected outcome:
Displayed table position matches physical table movement.

If the table moves but position feedback is inaccurate, remove the scanner from service. Incorrect table position can directly affect exam accuracy.

11. Perform a Controlled System Restart if Safe

Temporary software or communication faults can affect table control.

Action:

Expected outcome:
Table motion and position display return to normal if the issue was temporary.

If the issue clears after restart, perform a safe operational check and document the restart and successful table test.

12. Check for Room Power or System Power Issues

Table drive and control systems require stable power.

Action:

Expected outcome:
No external facility power issue is contributing to the table fault.

If power instability is suspected, involve facilities and vendor service as appropriate.

13. Review Error Messages and Event History

Table errors may be linked to position sensors, drive faults, collision detection, or communication issues.

Action:

Expected outcome:
The failure pattern is documented clearly for repair escalation.

If the Problem Persists

If table motion or positioning errors continue after checking patient safety, obstructions, emergency stops, controls, accessories, load conditions, table limits, power status, and restart behavior, the likely cause is internal to the table drive, lift system, position feedback system, encoder, interlock circuit, motion controller, or CT system communication pathway.

The device should be:

Do not continue patient scanning if table position is unreliable. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot CT table motion problems with a patient on the table unless the action is required to safely remove the patient. Unexpected movement, failure to stop, or incorrect positioning can create patient injury risk and scan accuracy problems. Move the patient to a backup device or alternate imaging pathway first so therapy and diagnostic care can continue safely.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CT staff reported the Philips Brilliance CT table would not advance into the gantry and displayed a table positioning error during exam setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found no patient obstruction, no activated emergency stop, and normal accessory setup, but table motion remained intermittent with inaccurate position feedback."

Resolution

What action was taken.

Example:
"Scanner was removed from service, labeled Out of Service, error details were documented, and Philips service was contacted for table drive and position feedback evaluation."

Helpful Details to Include (If Known)

Final Thought

CT table motion problems should be handled carefully because they affect both patient safety and exam accuracy. Start with external causes such as obstructions, controls, emergency stops, accessories, and load conditions before escalating. If motion is abnormal, noisy, unpredictable, or position feedback is inaccurate, remove the scanner from service and document the failure clearly.

That is successful troubleshooting.