Philips DigitalDiagnost C90

Acquisition Workspot Boot, Connectivity, or Image-Transfer Failure

On this page

Asset Type

Radiographic System

Manufacturer

Philips

Model

DigitalDiagnost C90

What This Guide Helps With

Troubleshooting Eleva workspot startup failure, loss of system communication, missing exams, or images not transferring to PACS.

Step-by-Step Troubleshooting

Ensure Patient Safety First

If the DigitalDiagnost C90 cannot reliably acquire, display, or transfer images, remove it from active patient imaging and notify Radiology staff.

Do not expose a patient when the acquisition workspot is unavailable, patient/exam data cannot be confirmed, or image transfer status is uncertain. Move the exam to another functioning radiographic system when clinically necessary.

Expected outcome: Patient imaging is performed only on equipment with reliable acquisition, review, and image-transfer capability.

Confirm the Exact Failure Condition

With the room removed from patient use, determine the reported symptom:

Record displayed messages, icons, destination names, timestamps, and whether the failure started after a power interruption, network outage, system update, or room shutdown.

Expected outcome: The issue is separated into boot, system communication, worklist/network, or image-transfer failure before troubleshooting continues.

Check Workspot Power and External Hardware

Inspect the workspot, monitor, keyboard/mouse, connected power cables, surge protection or approved power distribution components, and any visible external interface equipment.

Confirm:

The DigitalDiagnost C90 uses an Eleva workstation with solid-state drive-based operation for boot and access to examination information; a blank monitor or interrupted power path can appear to staff as a full acquisition failure.

Expected outcome: The workspot powers normally and displays the Philips application startup or login screen.

If resolved: Confirm normal startup and proceed only with approved functional verification before returning the room to service.

Rule Out a Display-Only Problem

If the workspot appears powered but the screen is blank or unreadable, verify the monitor power indicator, brightness controls if externally accessible, and securely seated external display connections.

Observe whether there are signs the workstation itself is running, such as indicator lights, cooling fan activity, login audio, or response to permitted input actions.

Expected outcome: A display issue is distinguished from a workstation boot failure.

If the monitor remains blank while the workstation appears powered: Remove the system from service and escalate for bench or vendor evaluation rather than repeatedly cycling power.

Perform One Controlled Restart When Appropriate

If no patient is involved, no active exposure is in progress, and there is no evidence of electrical damage or abnormal odor/heat, perform one controlled restart according to department-approved shutdown/startup practice.

Do not perform repeated hard power cycles or force shutdowns during suspected image-transfer activity, because examination or transfer status may be unclear afterward.

Observe and document:

Expected outcome: The acquisition workspot either restores normal operation or presents a repeatable failure condition suitable for escalation.

If resolved: Verify connectivity and image-transfer status before returning the system to clinical use.

Verify Visible System Communication Status

Once the workspot is operational, check for visible indications that required room components are available and communicating, such as generator, detector, tube head, table, or wall stand status as applicable to the installed room configuration.

The C90 uses the Eleva workspot as part of the examination workflow and may coordinate system operation and image-related activity through the workstation interface.

Check externally for:

Do not open cabinets or reseat internal communication boards.

Expected outcome: Required room components appear available and ready from the workspot.

If a room component remains unavailable: Remove the room from service and escalate for system communication evaluation.

Check Network Connectivity and Modality Worklist Function

Determine whether the failure includes network-dependent functions:

Coordinate with Radiology/PACS or IT support as appropriate to determine whether there is a known network, RIS, modality worklist, or PACS outage.

Inspect only accessible external network connections and documented room network indicators. Do not modify network settings, AE titles, IP configuration, DICOM destination settings, or security configuration without authorization.

Philips identifies the DigitalDiagnost C90 as supporting integration into hospital IT infrastructure and PACS.

Expected outcome: The issue is identified as either isolated to this radiographic room or associated with a broader network/PACS interruption.

If a broader outage is confirmed: Document the system condition and coordinate restoration through the appropriate support pathway before clinical use resumes.

Check Image-Transfer Queue and Local Image Availability

If acquisition is possible but images are not arriving in PACS, confirm whether the images remain visible locally on the workspot and whether a transfer queue or failed-send status is displayed.

Document:

Do not repeat exposures solely because an image has not transferred unless Radiology has confirmed that the original image is unusable or unavailable and repeat imaging is clinically required.

Expected outcome: Existing images are protected from unnecessary repeat exposure while the failure is isolated to local acquisition versus DICOM transfer.

If an authorized resend successfully delivers the image: Confirm receipt with Radiology/PACS and document the restored transfer.

Compare with Another Approved Destination or Room, When Available

Coordinate with Radiology/PACS support to determine whether:

Do not create test patients, alter destinations, or transmit patient information outside approved workflows.

Expected outcome: The failure is narrowed to the C90 workspot/system path or to an external PACS/network destination.

Complete a Safe Functional Verification

After any restored boot or connectivity condition, perform department-approved functional checks with no patient on the system.

Confirm:

Expected outcome: The acquisition workspot and image-transfer workflow are reliable for clinical operation.

If any failure repeats or verification cannot be completed: Keep the room out of service and escalate.

If the Problem Persists

If power, monitor status, accessible external connections, controlled restart, visible communication status, worklist availability, and approved image-transfer verification have been checked without resolution, common external causes have been ruled out.

The failure may involve the Eleva workspot computer, application software, internal communications, DICOM configuration, network interface, system controller, storage function, or another internal system component.

The device should be:

Do not continue repeated restarts, change network or DICOM configuration without authorization, or allow clinical imaging when image capture or transfer reliability is unresolved.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot acquisition, worklist, or image-transfer failures during an active patient examination. Move the patient to another available imaging system first, and confirm that images have reached the intended clinical destination before considering the exam complete.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiology reported that the Philips DigitalDiagnost C90 Eleva workspot powered on but would not send acquired images to PACS."

Cause

What was observed during troubleshooting.

Example:
"Workspot booted normally and images remained available locally, but the configured PACS destination showed failed transfer status after external connections and network availability were checked."

Resolution

What action was taken.

Example:
"Removed the room from clinical service, labeled it Out of Service, documented the affected exam-transfer status, and escalated to PACS/Philips support for image-transfer and system communication evaluation."

Helpful Details to Include (If Known)

Final Thought

A radiographic system is not dependable if the acquisition workspot cannot reliably start, identify examinations, communicate with room components, or deliver images for clinical review. Safe troubleshooting begins with protecting the patient from delayed imaging or unnecessary repeat exposure, continues with logical external checks, and ends with appropriate escalation and clear documentation when the failure is not safely correctable at the Clinical Engineering level.

That is successful troubleshooting.