Philips EPIQ Series

Image Dropout or Artifacts

On this page

Asset Type

Ultrasound System

Manufacturer

Philips

Model

EPIQ Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where the Philips EPIQ Series ultrasound system exhibits image dropout, streaking, shadowing, or other visual artifacts. Common presentations include:

These issues are often caused by external factors such as probe connection problems, cable integrity, or system settings before considering internal hardware failure.

Step-by-Step Troubleshooting

Confirm Proper Probe Connection

Inspect Probe and Cable for Damage

Swap with a Known Good Probe

Check Imaging Settings

Inspect System Cables and Power

Perform System Reboot

Run Built-in System Diagnostics

If the Problem Persists

After completing external checks, if artifacts continue:

Knowing when to escalate is proper Clinical Engineering troubleshooting. Do not attempt internal board-level repair.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ultrasound image shows black lines and streaking during exams.”

Cause

What was observed during troubleshooting.

Example:
“Probe cable shows intermittent signal; swapping with a known good probe resolved artifacts.”

Resolution

What action was taken.

Example:
“Defective probe removed from service and replaced. System tested with backup probe and imaging normal.”

Helpful Details to Include

Final Thought

Clinical Engineering troubleshooting relies on systematic evaluation, starting with external checks and user-accessible solutions before considering internal repairs. Proper documentation ensures patient safety and maintains system reliability. Escalation to repair should be done promptly if artifacts cannot be resolved externally.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide