Philips HeartStart MRx

Error: CO2 Equip Malfunction

On this page

Asset Type

Defibrillator

Manufacturer

Philips

Model

HeartStart MRx

What This Guide Helps With

This guide helps troubleshoot CO2 Equip Malfunction on the Philips HeartStart MRx when EtCO2 monitoring is unavailable, unreliable, or unable to initialize. This condition may affect CO2 waveform display, numeric EtCO2 values, FilterLine recognition, sampling flow, or CO2 module communication.

Because CO2 monitoring may be used to assess ventilation, airway status, and CPR effectiveness, this error should be treated as a patient-monitoring failure until verified and corrected.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check the FilterLine First

Inspect the CO2 Inlet and Sampling Connection

Inspect for Physical Damage or Fluid Exposure

Power Cycle and Retest

Confirm the Issue Is Specific to CO2

Review Logs or Operational Check Results

Retest With Known-Good Accessories

Stop When Resolved

If the Problem Persists

If the CO2 Equip Malfunction error remains after external checks, FilterLine replacement, inlet inspection, accessory seating verification, power cycling, and operational testing, common external causes have been ruled out.

Remove the Philips HeartStart MRx from service, label it Out of Service, and send it for repair or bench evaluation.

Persistent CO2 equipment malfunction may indicate an internal CO2 module, internal connection, sampling path, or communication-related failure that requires approved service evaluation.

Clinical Use Tip

Never troubleshoot CO2 monitoring on an active patient. Move the patient to a backup monitor or defibrillator first, and never rely on EtCO2 values from a device showing an unresolved CO2 equipment malfunction.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Philips HeartStart MRx reported CO2 Equip Malfunction. Staff stated CO2 would not initialize and no EtCO2 waveform or numeric value was available during monitoring setup."

Cause

What was observed during troubleshooting.

Example:
"CO2 error was confirmed during startup and CO2 selection. Known-good compatible FilterLine was tested, CO2 inlet was inspected for blockage or contamination, and the fault remained after external checks."

Resolution

What action was taken.

Example:
"Removed device from service and labeled Out of Service for bench evaluation. External CO2 accessory and sampling path checks were completed, and persistent CO2 Equip Malfunction was documented for repair."

Helpful Details to Include (If Known)

Final Thought

CO2 Equip Malfunction should be treated as a real monitoring failure until proven otherwise. Start with patient safety, then check the FilterLine, sampling path, connector area, power cycle behavior, and operational check results before escalating the device for repair.

That is successful troubleshooting.