Philips IntelliVue MP Series

Alarm: CHECK NURSE CALL RELAY

On this page

Asset Type

Patient Monitor

Manufacturer

Philips

Model

IntelliVue MP Series

What This Guide Helps With

This guide helps troubleshoot the CHECK NURSE CALL RELAY INOP on a Philips IntelliVue MP Series patient monitor. This alarm indicates the monitor is reporting a defective Nurse Call Relay board, which may prevent alarm notification from reaching the facility nurse call system even if local monitor alarms still appear to function.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Local Alarms Still Function

Check the Nurse Call Cable Connection

Inspect the Nurse Call Wall Jack or Interface Point

Swap With a Known-Good Nurse Call Cable if Available

Test the Monitor in a Known-Good Location if Practical

Check Configuration and Intended Use

Power Cycle and Recheck the INOP

Isolate the Failure

Stop When Resolved

If the Problem Persists

If CHECK NURSE CALL RELAY remains present after external checks, cable swap, location/interface check, configuration review, and power cycle, the external causes have been ruled out. Remove the monitor from service, label it Out of Service, and send it for repair or bench evaluation.

A persistent CHECK NURSE CALL RELAY INOP strongly indicates failure of the Nurse Call Relay I/O board. Replacement of the Nurse Call Relay I/O board is the appropriate service action when the fault follows the monitor and does not clear after external troubleshooting.

Clinical Use Tip

Never troubleshoot nurse call relay output on an active patient when that monitor is being relied on for alarm notification. Move the patient to a backup monitor or ensure clinical staff maintain direct bedside monitoring and local alarm awareness first. Maintain therapy and monitoring continuity before testing cables, configuration, room interfaces, or relay output behavior.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Philips IntelliVue MP Series monitor displaying CHECK NURSE CALL RELAY INOP. Nurse call relay output may not be reliable for alarm notification."

Cause

What was observed during troubleshooting.

Example:
"External nurse call cable, wall interface, and configuration checked. INOP followed the monitor. Persistent alarm indicates likely defective Nurse Call Relay I/O board."

Resolution

What action was taken.

Example:
"Removed monitor from service and labeled Out of Service. Sent for repair or bench evaluation for Nurse Call Relay I/O board replacement and post-repair nurse call verification."

Helpful Details to Include (If Known)

Final Thought

Nurse call relay alarms require careful patient safety awareness because the monitor may still appear functional locally while external alarm notification is compromised. Start with simple external checks, verify whether the issue follows the monitor, and escalate when the system-reported INOP persists. Clear CCR documentation protects patient safety and supports the repair process.

That is successful troubleshooting.