Philips IntelliVue MP Series

Alarm: MsmtSrv not Supp

On this page

Asset Type

Patient Monitor

Manufacturer

Philips

Model

IntelliVue MP Series

What This Guide Helps With

This guide helps troubleshoot the MsmtSrv not Supp INOP alarm on a Philips IntelliVue MP Series patient monitor when the monitor does not support the connected measurement server configuration. This may prevent the connected measurement server from being recognized and may make required parameters unavailable.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check the Measurement Server Connection

Power Cycle the Monitor With the Measurement Server Connected

Test With a Known-Compatible Measurement Server

Check for Too Many Connected Measurement Servers

Verify the Measurement Server Type Is Supported

Check Software Revision Compatibility

Compare Against a Known-Good Monitor

Correct the Configuration if the Cause Is Identified

If the Problem Persists

If the MsmtSrv not Supp INOP remains after checking connections, reducing connected measurement servers, testing known-compatible servers, and verifying software compatibility, external and common configuration causes have been ruled out.

A persistent alarm may indicate an incompatible monitor software revision, incompatible measurement server software revision, unsupported measurement server type, incorrect system configuration, or a monitor or measurement server communication fault requiring service evaluation.

Remove the monitor from service, label it Out of Service, and send it for bench evaluation or Philips/service-level repair support.

Clinical Use Tip

Never troubleshoot measurement server compatibility on an active patient if required parameters are missing. Move the patient to a backup device first, then evaluate the monitor and measurement server combination safely while maintaining therapy and monitoring continuity.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Philips IntelliVue MP Series monitor displays MsmtSrv not Supp INOP when measurement server is connected. Required parameters may not be available."

Cause

What was observed during troubleshooting.

Example:
"Connected measurement server configuration was not supported. Possible wrong software revision, unsupported measurement server type, or too many measurement servers connected."

Resolution

What action was taken.

Example:
"Verified INOP, removed monitor from patient use, inspected and reseated measurement server connection, power cycled monitor, tested with known-compatible measurement server, and checked for unsupported or extra measurement servers. Device returned to service after compatible configuration was confirmed, or removed from service for software compatibility review/bench evaluation if INOP persisted."

Helpful Details to Include (If Known)

Final Thought

The MsmtSrv not Supp INOP should be approached as a compatibility and configuration problem first, not an automatic hardware failure. Clinical Engineering should verify the alarm, protect patient monitoring, check the connected measurement server setup, confirm supported combinations, and escalate when software or system compatibility cannot be corrected at the user-accessible level.

That is successful troubleshooting.