On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoot Lumify compatibility messages involving the mobile device, operating system, application version, transducer, permissions, or recent software changes.
Step-by-Step Troubleshooting
1. Protect the Patient and Use a Verified Alternative
Do not bypass a compatibility warning or use an unsupported imaging configuration for patient care. Move the exam to a verified Lumify configuration or another ultrasound system.
Expected outcome: Clinical imaging proceeds on equipment with a known supported configuration.
2. Capture the Exact Compatibility Message
Record the complete message shown by Lumify or the mobile device without paraphrasing it.
Determine whether the warning references the mobile device, operating system, connected hardware, application, or another condition.
Expected outcome: The exact reported compatibility condition is documented before any changes are made.
3. Confirm the Correct Mobile Device
Verify that the mobile device presented for troubleshooting is the one normally assigned or approved for the Lumify system.
Check its model using normal device settings.
Expected outcome: The mobile device identity is confirmed. If an incorrect or unapproved device was substituted, restore the approved device and verify operation.
4. Verify the Operating-System Version
Review the current operating-system version through normal device settings and determine whether it recently changed.
Do not downgrade, root, jailbreak, or otherwise modify the operating system to bypass compatibility controls.
Expected outcome: The installed operating-system version is accurately identified for comparison with approved Philips compatibility information.
5. Verify the Lumify Application Version
Confirm the Lumify application version and whether it was recently installed or updated.
Document any mismatch between organizationally approved software and the installed application.
Expected outcome: The current application version is known and can be evaluated against the approved environment.
6. Verify the Connected Transducer and Interface
Confirm that the intended Lumify transducer is being used and that any required approved interface or adapter is correctly connected.
Inspect the connector for damage or incomplete seating.
Expected outcome: The expected transducer and external connection hardware are present and correctly connected.
7. Restart the Approved Configuration
After verifying the hardware and software environment, perform a normal restart of the mobile device and relaunch Lumify.
Reconnect the transducer after the device has stabilized.
Expected outcome: The compatibility warning clears and normal transducer detection returns. If so, proceed to functional verification.
8. Compare With a Known-Good Supported Setup
When available, test the same transducer with another known-good supported mobile device or compare the affected device with another Lumify system configured according to current organizational standards.
Expected outcome: The comparison helps determine whether the condition follows the mobile device/software environment or the transducer.
9. Verify Normal Imaging After Correction
If the compatibility condition is corrected through an approved configuration, perform a nonpatient test scan and confirm transducer recognition and application stability.
Expected outcome: Lumify operates normally in a verified supported configuration. Troubleshooting can stop.
10. Escalate Unresolved Compatibility Problems
If the device still reports unsupported hardware or software after identifying the exact versions and confirming the approved configuration, do not attempt workarounds.
Expected outcome: The affected configuration is withheld from clinical use and escalated to qualified support.
If the Problem Persists
External configuration checks have been completed. The remaining cause may involve an unsupported software combination, device-specific compatibility issue, application defect, required vendor update, or other service-level configuration problem.
Remove the affected system from service, label it Out of Service when it cannot be used in an approved configuration, and send it for bench or software evaluation. Use current Philips documentation and organizational mobile-device management procedures. Configuration changes should be performed only by qualified personnel using supported methods.
Before return to service, verify the hardware and software combination is approved and perform full Lumify functional testing. Avoiding unsupported workarounds is part of proper troubleshooting.
Clinical Use Tip
A device that physically connects to a Lumify transducer is not automatically a supported clinical configuration.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that Lumify displayed an unsupported-device message and would not allow normal scanning."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found that the mobile device operating system had been updated outside the currently approved Lumify configuration."
Resolution
What action was taken.
Example:
"Clinical Engineering removed the affected device from use and restored service using a verified compatible mobile device, then confirmed normal Lumify imaging operation."
Helpful Details to Include (If Known)
- Exact compatibility message
- Mobile device model
- Operating-system version
- Lumify application version
- Transducer used
- Adapter or connection hardware
- Recent software or device changes
- Known-good compatible setup tested
- Whether restart changed the condition
- Final approved configuration and status
Final Thought
Compatibility problems should be verified against the complete hardware and software environment, not bypassed. Use approved configurations, document exact versions, and escalate unresolved compatibility conditions appropriately.
That is successful troubleshooting.