What This Guide Helps With
Troubleshoots detector, positioning, exposure, environmental, communication, and calibration-related image artifacts before assuming permanent detector failure.
Step-by-Step Troubleshooting
1. Protect the Patient and Avoid Repeated Diagnostic Exposures
Do not repeatedly expose a patient while determining whether an image-quality problem is reproducible.
Use approved test objects or another verified imaging system as appropriate.
Expected outcome: Artifact troubleshooting proceeds without unnecessary patient radiation exposure.
2. Confirm the Artifact Pattern
Determine whether the image shows:
- Fixed lines
- Bands
- Excessive noise
- Residual or lag image
- Nonuniform shading
- Repeating detector pattern
- Intermittent corruption
Save representative images when permitted for service documentation.
Expected outcome: The artifact is clearly characterized and can be compared between tests.
3. Determine Whether the Problem Is Detector-Specific
Identify which detector was used and whether the artifact occurs with every examination.
If another approved compatible detector is available, compare images under controlled conditions.
Expected outcome: The problem is narrowed to one detector or shown to affect the system more broadly.
4. Inspect the Detector Externally
Check the detector for:
- Impact damage
- Flexing or housing damage
- Contamination
- Liquid exposure
- Damaged battery or accessible connectors
- Foreign material affecting positioning
Expected outcome: No visible physical condition explains the artifact. A damaged detector should be removed from service.
5. Verify Detector Charge and Communication Stability
Confirm the detector is adequately charged and maintains stable communication during acquisition and image transfer.
Expected outcome: Image artifacts do not correlate with low detector charge or communication interruption.
6. Check Exposure and Positioning Conditions
Confirm the test image is obtained with appropriate:
- Detector orientation
- Tube alignment
- Source-to-receptor geometry
- Exposure selection
- Object placement
Avoid diagnosing detector failure from a poorly controlled image.
Expected outcome: A properly controlled acquisition still reproduces the artifact before further escalation.
7. Perform Approved Detector Correction or Calibration Workflow
If authorized Clinical Engineering procedures permit, perform only the manufacturer-approved normal detector correction or calibration process.
Do not use restricted service procedures without authorization.
Expected outcome: The detector completes the approved correction process and subsequent test images are uniform. If the artifact disappears and remains absent, proceed to final verification.
8. Compare Multiple Test Images
Acquire several approved test images to determine whether the defect:
- Remains in the same location
- Changes with positioning
- Appears only after prior exposures
- Is intermittent
Expected outcome: A consistent pattern provides evidence for detector-specific or system-level escalation.
9. Perform Final Image-Quality Verification
Verify:
- No persistent lines or bands remain
- Uniform areas appear acceptable
- No obvious lag is present
- Noise is not abnormal for the controlled test
- Detector communication remains stable
Use approved image-quality procedures and test equipment where applicable.
Expected outcome: Image quality is repeatable and acceptable under approved verification procedures. Troubleshooting can stop.
10. Escalate Persistent Image Artifacts
If artifacts persist after controlled acquisition, external inspection, known-good comparison, and approved correction procedures, remove the affected detector or system from service.
Expected outcome: Unreliable diagnostic image quality is not returned to clinical use.
If the Problem Persists
Common external causes such as detector condition, charge, communication, positioning, acquisition setup, and approved correction procedures have been ruled out.
The remaining problem may involve detector electronics, detector calibration data, image-processing functions, communication hardware, or another service-level condition.
The affected equipment should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate Philips documentation and approved image-quality test equipment
- Repaired or configured only by qualified personnel
Complete appropriate image-quality and detector performance verification before return to service.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
A detector that produces a repeatable image defect in the same location should not remain in diagnostic use simply because the image is still viewable.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Radiology reported repeated vertical lines on images acquired with one MobileDiagnost wDR detector."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering confirmed the artifact followed the same detector and found the detector required an approved correction procedure."
Resolution
What action was taken.
Example:
"Clinical Engineering completed the authorized detector correction process and verified multiple test images were free of the reported line artifact."
Helpful Details to Include (If Known)
- Artifact type
- Detector identification
- Whether artifact follows one detector
- Detector physical condition
- Battery status
- Exposure and positioning conditions
- Known-good comparison result
- Correction or calibration performed
- Before-and-after test images
- Final device status
Final Thought
Image-quality troubleshooting should be controlled and repeatable. Protect the patient from unnecessary exposure, determine whether the problem follows the detector, and rule out positioning, charge, communication, and approved calibration causes before escalating for repair.
That is successful troubleshooting.