Philips PageWriter TC70

Error: Printer temperature is too high

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

Philips

Model

PageWriter TC70

What This Guide Helps With

This guide helps troubleshoot the Philips PageWriter TC70 error Printer temperature is too high. This condition can stop or limit ECG report printing, rhythm strip printing, or printer operation when the print head temperature is above the expected operating range.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Stop Printing and Power Down the Cardiograph

Check the Environment Around the Unit

Inspect the Printer Area Externally

Verify Correct Paper Loading

Restart the Cardiograph

Perform a Controlled Test Print

Check for Repeated Heavy Printing Conditions

Stop When Resolved

If the Problem Persists

If Printer temperature is too high returns immediately after cooldown, appears at startup, or returns during normal printing, external and common causes have been ruled out.

A persistent printer temperature error may indicate a printer assembly, print head temperature sensing issue, printer control issue, or internal printer-related failure.

Remove the PageWriter TC70 from service.

Label it Out of Service.

Send it for repair or bench evaluation. If required by facility process, contact the Philips Response Center for further support.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot printer temperature errors while the device is needed for an active patient ECG. Move the patient to a backup device first and use another verified ECG machine if printed reports are required for patient care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Philips PageWriter TC70 displayed Printer temperature is too high. Staff reported printing was unavailable after ECG or rhythm printing."

Cause

What was observed during troubleshooting.

Example:
"Print head over-temperature condition observed. Printer area inspected for paper jam, paper loading issue, restricted paper path, and environmental heat exposure. Error cleared after shutdown and 20-minute cooldown."

Resolution

What action was taken.

Example:
"Powered down cardiograph for 20 minutes, inspected printer drawer and paper path, verified paper was loaded correctly, restarted unit, and completed successful test print. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

Printer temperature errors should be handled calmly and logically. Protect patient care first, let the print head cool, check simple external causes, verify printer operation, and escalate when the error persists after basic troubleshooting. Clear documentation helps show whether the issue was temporary overheating or a likely internal printer failure.

That is successful troubleshooting.