Philips PageWriter TC70

Error: Printer underflow

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

Philips

Model

PageWriter TC70

What This Guide Helps With

This guide helps troubleshoot Printer underflow errors on the Philips PageWriter TC70. This condition may appear during rhythm printing and can be associated with intermittent Patient Interface Module communication, signal acquisition interruptions, or patient cable connection problems. The issue may interrupt rhythm strip printing, create blank or delayed print sections, or occur alongside ECG signal instability.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check Whether Printing Stops, Pauses, or Produces Poor Output

Check AC Power and Battery Status

Inspect the Printer Area for Simple Issues

Check the PIM Connection

Inspect the Patient Data Cable

Inspect Lead Wires and Patient Cable Accessories

Run or Review PIM Test Results

Try a Known-Good PIM

Check for a Device-Specific Pattern

Restart and Recheck

Stop When Resolved

If the Problem Persists

If the Printer underflow error continues after checking AC power, battery status, printer paper path, PIM connection, patient data cable, lead wires, known-good accessories, and PIM test results, external and common causes have been ruled out.

Persistent Printer underflow errors may indicate an unresolved PIM communication fault, accessory interface issue, or internal cardiograph communication problem.

Remove the Philips PageWriter TC70 from service, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot intermittent PIM communication errors on an active patient. Move the patient to a backup device first, then test cables, PIM connections, accessories, and rhythm printing behavior in a controlled setting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Philips PageWriter TC70 displayed Printer underflow during rhythm printing. Staff reported interrupted rhythm strip printing and intermittent ECG signal acquisition behavior."

Cause

What was observed during troubleshooting.

Example:
"Printer underflow condition confirmed during rhythm printing. PIM communication path inspected. Patient data cable and PIM connection checked due to intermittent acquisition communication behavior."

Resolution

What action was taken.

Example:
"Removed device from patient use. Checked AC power, printer paper path, PIM connection, patient data cable, and lead wires. Reseated connections and repeated rhythm print testing. Device passed functional testing and was returned to service, or labeled Out of Service for bench evaluation if unresolved."

Helpful Details to Include (If Known)

Final Thought

Printer underflow on the Philips PageWriter TC70 should be approached as a communication-related acquisition problem first, not just a printer problem. Patient safety comes first, followed by logical external checks of power, printer condition, PIM connection, patient data cable, lead wires, and PIM test results. If the error persists after those checks, escalation for repair is appropriate and should be clearly documented.

That is successful troubleshooting.