On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting pacer activation, output, sensing, electrode-connection, intermittent pacing, and simulated capture failures caused by power, cables, electrodes, ECG setup, or controls.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not perform extended troubleshooting while the LIFEPAK 12 is actively supporting a patient who requires pacing.
If pacing fails during patient care:
- Notify the clinical team immediately.
- Obtain another verified monitor/defibrillator or alternate pacing method.
- Continue continuous ECG and hemodynamic assessment.
- Do not delay clinically required therapy while repeatedly reconnecting accessories or cycling power.
- Transfer pacing support to the replacement device as soon as safely possible.
Expected outcome: Patient care no longer depends on a device with uncertain pacer output.
Continue Clinical Engineering troubleshooting only after the LIFEPAK 12 has been removed from active patient use.
2. Identify the Exact Pacing Failure
In a controlled, nonclinical area, document whether:
- Pressing PACER produces no response.
- The PACER indicator illuminates, but current will not increase.
- Current increases, but no pacing markers appear.
- Pacing begins and then stops.
- CONNECT ELECTRODES, PACING STOPPED, CURRENT FAULT, or PACER FAULT appears.
- The displayed pacing rate does not match the measured output.
- Output is present on a simulator, but capture was not achieved clinically.
Expected outcome: The complaint is narrowed to power, connection, control, sensing, output, or clinical-capture behavior.
3. Check Device Power and Battery Condition
Confirm that the LIFEPAK 12:
- Powers on normally.
- Completes startup without a service indication.
- Has adequately charged batteries installed.
- Operates from an approved auxiliary power source when available.
- Does not restart, dim, or shut down when the pacer is activated.
The manufacturer lists power-off and low battery conditions as possible reasons the device may not function when PACER is pressed.
Expected outcome: The unit remains powered and stable when pacing mode is selected.
If replacing the battery or using verified external power resolves the issue, complete functional testing and stop.
4. Inspect the Therapy Cable and Connections
Inspect the QUIK-COMBO therapy cable for:
- Bent, recessed, contaminated, or damaged connector pins.
- Cracked insulation or exposed conductors.
- Loose connector housings.
- Severe cable kinks or strain damage.
- Fluid intrusion or residue.
- An incomplete connection at the LIFEPAK 12 or test load.
Disconnect and firmly reconnect the therapy cable. When available, repeat the test with a verified compatible therapy cable.
A disconnected therapy cable or electrode connection can prevent current adjustment and produce CONNECT ELECTRODES or PACING STOPPED messages.
Expected outcome: The therapy connection is secure, and no connection-related message remains.
If a known-good cable corrects the problem, replace the defective cable and stop.
5. Check the Therapy Electrodes or Simulator Connection
For bench evaluation, connect the LIFEPAK 12 to a verified defibrillator/pacer analyzer using an approved test load or adapter.
For a reported clinical event, review whether the electrodes were:
- Compatible QUIK-COMBO electrodes.
- Within their expiration date.
- Fully connected.
- Properly adhered to prepared skin.
- Damaged, dried, contaminated, or previously used.
- Positioned correctly for pacing.
The LIFEPAK 12 may display CONNECT ELECTRODES when electrodes are disconnected, poorly adhered, or outdated. After reconnection, pacing current remains at 0 mA until it is manually increased again.
Expected outcome: The analyzer or electrodes are recognized, and pacing current can be selected.
If replacing the electrodes or correcting the connection resolves the issue, stop.
6. Confirm Separate ECG Monitoring Connections
Verify that the patient ECG cable and monitoring electrodes are connected when evaluating demand pacing.
The LIFEPAK 12 requires separate ECG electrodes and a patient ECG cable during pacing; the therapy electrodes cannot simultaneously provide pacing current and the required ECG monitoring signal.
Check for:
- ECG LEADS OFF messages.
- Loose or dried monitoring electrodes.
- Damaged lead wires.
- Poor skin preparation.
- ECG electrodes positioned too close to the therapy electrodes.
Expected outcome: A stable ECG waveform is displayed with recognizable QRS complexes.
7. Verify Pacer Activation and Control Response
Press PACER and confirm:
- The PACER indicator illuminates.
- The pacing rate can be adjusted.
- The pacing current can be increased from 0 mA.
- Pacing markers appear on the displayed waveform or analyzer.
- The PACER indicator responds with each delivered pulse.
The CURRENT button adjusts output in 10 mA increments, while the selector adjusts it in 5 mA increments.
Expected outcome: Rate and current controls respond normally, and pacing pulses are displayed or measured.
If the PACER indicator illuminates but current cannot be increased, recheck the therapy cable and load before suspecting an internal failure.
8. Check Demand-Pacing Sensing
In demand mode, confirm that sense markers appear near the middle of each QRS complex.
If pacing markers are absent despite current being above 0 mA:
- Confirm the selected pacing rate is above the simulated intrinsic rate when testing.
- Reduce excessive ECG size if artifact or T-waves are being oversensed.
- Establish a cleaner ECG signal.
- Select another ECG lead.
- Move ECG electrodes farther from the pacing electrodes.
- Remove nearby radio-frequency interference.
The manufacturer identifies low pacing rate, oversensing, ECG artifact, excessive ECG size, poor electrode positioning, and radio interference as possible causes of absent or unstable pacing output.
Expected outcome: The LIFEPAK 12 properly senses intrinsic QRS complexes and delivers pacing pulses when required.
9. Test Pacer Output With an Analyzer
Connect the unit to a calibrated defibrillator/pacer analyzer and verify:
- Selected pacing rate against measured rate.
- Selected current against measured current.
- Output stability over several minutes.
- Operation at multiple rate and current settings.
- Demand-mode inhibition and triggering.
- Nondemand output when appropriately selected for bench testing.
- No intermittent loss of output when the cable is gently repositioned.
Do not determine pacing output by touching electrodes together or attempting an uncontrolled discharge.
Expected outcome: Measured pacing rate and current remain within the analyzer and manufacturer-approved performance limits.
If analyzer results are acceptable, investigate accessories, electrode application, ECG sensing, or the distinction between electrical and mechanical capture.
10. Distinguish Device Output From Clinical Capture
Pacing markers or analyzer output confirm electrical pulse delivery but do not prove mechanical capture in a patient.
During clinical use, capture must be evaluated by qualified clinical staff using:
- ECG evidence of electrical capture.
- Palpable pulse.
- Blood-pressure response.
- Other appropriate hemodynamic monitoring.
Failure to capture may result from insufficient selected current or patient-specific factors rather than equipment failure. The manufacturer directs users to increase pacing current until electrical capture occurs and then confirm mechanical capture by pulse or blood pressure.
Expected outcome: A reported “no capture” complaint is correctly separated from a true pacer-output malfunction.
11. Evaluate Intermittent Pacing Stops
If pacing stops unexpectedly, check whether:
- PACER was inadvertently pressed.
- ENERGY SELECT or CHARGE was pressed, which stops pacing.
- A therapy electrode or cable became disconnected.
- Radio equipment was operating near the device.
- A service indicator appeared.
- PACER FAULT or CURRENT FAULT was displayed.
Cycle power once in a controlled setting and repeat the analyzer test. Do not repeatedly return a device to service after unexplained therapy interruptions.
The manufacturer associates spontaneous pacing stops with disconnected electrodes, defibrillation controls being activated, radio-frequency interference, or internal faults.
Expected outcome: The interruption is either reproduced and traced to an external cause or classified as an internal fault.
12. Complete a Final Functional Check
After correcting an external cause:
- Repeat pacing tests at several rates and current levels.
- Confirm the displayed settings agree with analyzer measurements.
- Disconnect the therapy cable and verify the expected alarm and connection message.
- Reconnect the cable and confirm the alarm clears.
- Verify that current remains at 0 mA until manually increased.
- Inspect the unit for service indicators.
- Complete any required electrical-safety and performance-verification procedures.
Expected outcome: Pacer operation is stable, repeatable, and ready for clinical use.
If the Problem Persists
If verified batteries, power, ECG leads, therapy cable, electrodes, analyzer connections, sensing settings, and operating controls have been ruled out, the problem is likely internal.
Remove the LIFEPAK 12 from service if:
- Pacing current cannot be generated or adjusted.
- Measured rate or current is outside acceptable tolerance.
- Output is intermittent.
- PACER FAULT, CURRENT FAULT, or a service indication recurs.
- The PACER control or selector does not respond reliably.
- Pacing stops without an identifiable external cause.
The device should be:
- Removed from service.
- Labeled Out of Service.
- Sent for qualified repair or bench evaluation.
- Performance-tested before being returned to clinical use.
Knowing when to stop and escalate a therapy-function failure is proper troubleshooting.
Clinical Use Tip
Never troubleshoot questionable pacing output on an active patient. Move the patient to another verified device or alternate pacing method first, then evaluate the LIFEPAK 12 under controlled conditions with a calibrated analyzer.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the LIFEPAK 12 entered pacing mode, but pacing current could not be increased and CONNECT ELECTRODES remained displayed."
Cause
What was observed during troubleshooting.
Example:
"Inspection found an intermittent open connection in the QUIK-COMBO therapy cable near the device connector."
Resolution
What action was taken.
Example:
"Replaced the therapy cable and verified pacing rate, current output, demand sensing, disconnect alarms, and stable operation using a calibrated defibrillator/pacer analyzer."
Helpful Details to Include (If Known)
- Battery condition and auxiliary power tested
- Startup or service indicators observed
- Exact displayed messages
- Therapy cable inspected or swapped
- Electrode type and expiration date
- ECG lead and waveform quality
- Demand or nondemand mode
- Selected and measured pacing rate
- Selected and measured pacing current
- Analyzer model and test results
- Intermittent behavior reproduced
- Unusual heat, smell, noise, or physical damage
- Final device status
Final Thought
External pacing is a critical therapy function. Patient safety requires immediate backup coverage, logical separation of accessories and sensing problems from true output failure, appropriate analyzer testing, and prompt escalation when reliable operation cannot be confirmed. Clear CCR documentation preserves the clinical complaint, technical findings, and final disposition.
That is successful troubleshooting.