On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide addresses situations where the LIFEPAK 12:
- Fails to print ECG strips or event documentation
- Produces incomplete, garbled, or faint printouts
- Shows printer error messages or warnings
Focus is on external, logical troubleshooting first, avoiding internal printer repair.
Step-by-Step Troubleshooting
Verify Printer Power
- Ensure the printer module or integrated printer has power.
- Check any power switches or LEDs.
- Why: Without power, no printouts are possible.
Check Paper Supply and Orientation
- Confirm there is paper in the printer tray or roll.
- Ensure it is installed correctly (not upside-down or jammed).
- Why: Improper paper loading is a common cause of print failures.
Inspect Paper Path for Obstructions
- Remove any torn paper, debris, or foreign objects.
- Gently clean the paper feed path if needed.
- Why: Obstructions can prevent printing or cause misfeeds.
Confirm Printer Connection / Module Seating
- Ensure the printer module is properly seated if removable.
- Check that all connectors and latches are secure.
- Why: Loose connections may prevent data transfer to the printer.
Check Device Settings
- Verify print settings in the LIFEPAK 12 menu.
- Ensure event documentation printing is enabled.
- Why: Incorrect configuration may suppress printing.
Test with a Known Good Paper Roll
- Swap the current paper with a fresh, compatible roll.
- Why: Paper type or quality issues can prevent clear printouts.
Perform a Test Print / Self-Test
- Use the printer’s self-test function if available.
- Confirm if the printer prints independently of patient data.
- Why: Isolates printer hardware vs device software issues.
If the Problem Persists
External checks (power, paper, connections, settings) have been ruled out. The issue is likely internal (printer mechanism or board-level fault).
- Remove the device from service
- Label it Out of Service
- Send for repair or bench evaluation
Knowing when to escalate prevents further patient risk and wasted time.
Clinical Use Tip
Do not attempt to print patient data during active resuscitation if the printer is unreliable. Document manually in the patient chart and resume device-based documentation only when the printer is verified operational.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
“Printer will not print ECG strips or event logs.”
Cause
What was observed during troubleshooting.
Example:
“Paper loaded correctly, printer powered, connections verified, but print test failed.”
Resolution
What action was taken.
Example:
“Device removed from service, labeled Out of Service, sent to Biomed for repair.”
Helpful Details to Include
- Printer LEDs (status indicators) observed
- Paper type and orientation used
- Any unusual sounds from printer
- Last successful print timestamp
- Error messages displayed on device screen
Final Thought
Ensuring accurate documentation is critical for both patient care and regulatory compliance. Logical troubleshooting that starts with power, paper, and connections saves time and prevents unnecessary device downtime. When external checks fail, escalation is proper and safe.
That is successful troubleshooting.