Physio-Control LIFEPAK 15

Error: PA_ERROR_PULSE_WIDTH Pacing Pulse Width Error

On this page

Asset Type

Defibrillator

Manufacturer

Physio-Control

Model

LIFEPAK 15

What This Guide Helps With

Troubleshooting PA_ERROR_PULSE_WIDTH, where pacing pulse duration is too long and may affect safe pacer therapy delivery.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Remove the Device From Clinical Use

Check External Power and Basic Device Condition

Inspect Therapy Cable and Pacing Accessories

Review the Service Log

Attempt a Controlled Bench Recheck Only

Verify Pacer Characteristics Before Return to Use

Stop When Resolved

If the Problem Persists

If PA_ERROR_PULSE_WIDTH or therapy service code a20e persists after external power, accessories, physical condition, and basic operational checks are ruled out, the fault should be treated as an internal therapy/pacing subsystem failure.

Remove the LIFEPAK 15 from service, label it Out of Service, and send it for repair or bench evaluation. Do not return the unit to clinical use until corrective action is completed and pacer characteristics are verified.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot on an active patient. If pacing therapy is needed, move the patient to a backup device first and have staff immediately use another verified defibrillator/pacer. A pacing pulse width fault can affect therapy delivery, so maintain therapy continuity and treat this as a patient safety issue.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported LIFEPAK 15 displaying PA_ERROR_PULSE_WIDTH with therapy service code a20e. Service information indicates pacing pulse too long."

Cause

What was observed during troubleshooting.

Example:
"External power source, therapy cable, pacing electrodes, and device condition checked. Persistent therapy service code indicates likely internal pacing/therapy subsystem fault requiring repair evaluation."

Resolution

What action was taken.

Example:
"Removed defibrillator from service and labeled Out of Service. Processed service log and routed unit for corrective repair. Pacer characteristics to be verified with approved analyzer before return to use."

Helpful Details to Include (If Known)

Final Thought

A pacing pulse width error on a defibrillator is not a minor nuisance alarm. Clinical Engineering should confirm the code, rule out external causes, protect patient safety, and escalate appropriately when the therapy subsystem remains suspect. Clear CCR documentation supports both safe service decisions and future repair history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide