Physio-Control LIFEPAK 15

Error: TH_ERROR_PA_CURRENT_OUT_OF_RANGE Pacing Current Out of Range

On this page

Asset Type

Defibrillator

Manufacturer

Physio-Control

Model

LIFEPAK 15

What This Guide Helps With

This guide helps troubleshoot TH_ERROR_PA_CURRENT_OUT_OF_RANGE on the Physio-Control LIFEPAK 15 when the selected pacing current does not match the actual delivered pacing output tolerance. This is a therapy-output accuracy concern and should be handled as a pacing fault, not as a routine user setup issue.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Translate the Error Into the Real-World Failure

Check for Obvious External Issues

Check the Therapy Cable and Accessories

Verify Power Condition

Review the Service Log or Reported Service Code

Do Not Attempt Clinical Return Based Only on Restarting

Determine Whether the Issue Is Accessory-Related or Device-Related

Stop When Resolved

If the Problem Persists

If external accessories, cable condition, power condition, and setup have been ruled out, a persistent TH_ERROR_PA_CURRENT_OUT_OF_RANGE with therapy service code a011 strongly indicates an internal pacing therapy failure.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Do not return the LIFEPAK 15 to service until pacing output accuracy and therapy function are verified according to facility and manufacturer service requirements.

Clinical Use Tip

Never troubleshoot pacing output problems on an active patient. If pacing is needed and the LIFEPAK 15 reports a pacing current tolerance error, move the patient to a backup verified defibrillator/monitor first and maintain therapy continuity before equipment troubleshooting continues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"LIFEPAK 15 reported TH_ERROR_PA_CURRENT_OUT_OF_RANGE with therapy service code a011. Reported condition indicates pacing current out of tolerance between selected current and actual current."

Cause

What was observed during troubleshooting.

Example:
"External cable, accessory, and power checks were performed. Error persisted with no correctable external issue found. Fault indicates likely internal pacing output tolerance failure."

Resolution

What action was taken.

Example:
"Removed LIFEPAK 15 from service and labeled Out of Service. Sent for repair or bench evaluation. Pacing output verification required after repair before return to clinical use."

Helpful Details to Include (If Known)

Final Thought

A pacing current out-of-range error on a defibrillator is a therapy accuracy concern, not a routine user issue. Clinical Engineering should confirm the error, rule out simple external causes, remove the device from service when the condition persists, and document the failure clearly. Proper escalation protects the patient and ensures the unit is not returned until pacing output is verified.

That is successful troubleshooting.